BACKGROUND

Impact of contact force sensing (CFS) on ablation of persistent atrial fibrillation (PeAF) is unknown.

OBJECTIVE

The purpose of the TOUCH AF (Therapeutic Outcomes Using Contact force Handling during Ablation of Persistent Atrial Fibrillation) randomized trial was to compare CFS-guided ablation to a CFS-blinded strategy.

METHODS

Patients (n = 128) undergoing first-time ablation for persistent AF were randomized to a CFS-guided vs CFS-blinded strategy. In the CFS-guided procedure, operators visualized real-time force data. In the blinded procedure, force data were hidden. Wide antral pulmonary vein isolation plus a roof line were performed. Patients were followed at 3, 6, 9, and 12 months with clinical visit, ECG, and 48-hour Holter monitoring. The primary endpoint was cumulative radiofrequency (RF) time for all procedures. Atrial arrhythmia >30 seconds after 3 months was a recurrence.

RESULTS

PeAF was continuous for 26 weeks (interquartile range [IQR] 13-52), and left atrial size was 45 ± 5 mm. Force in the CFS-blinded and CFS-guided arms was 12 g [IQR 6-20] and 14 g [IQR 9-20] (P = .10), respectively. Total RF time did not differ between CFS-guided and CFS-blinded groups (49 ± 14 min vs 50 ± 20 min, respectively; P = .70). Single procedure freedom from atrial arrhythmia was 60% in the CFS-guided arm and 63% in the CFS-blinded arm off drugs. Lesions with gaps were associated with significantly less force (11.4 g [IQR 6-19] vs 13.2 g [IQR 8-20], respectively; P = .0007) and less force-time integral (174 gs [IQR 91-330] vs 210 gs [IQR 113-388], respectively; P <.001).

CONCLUSION

CFS-guided ablation resulted in no difference to RF time or 12-month outcome. Lower force/force-time integral was associated with significantly more gaps.

Для цитирования: Conti Sergio, Weerasooriya Rukshen, Novak Paul, Champagne Jean, Lim Hong Euy, Macle Laurent, Khaykin Yaariv, Pantano Alfredo, Verma Atul. Contact force sensing for ablation of persistent atrial fibrillation: A randomized, multicenter trial.. Heart rhythm. 2017. DOI: 10.1016/j.hrthm.2017.10.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.