DOI 10.3390/bioengineering11040382

Медицинская библиотека · Bioengineering (Basel, Switzerland) · 2024

Limb Volume Measurements: A Comparison of Circumferential Techniques and Optoelectronic Systems against Water Displacement.

Farina Giovanni, Galli Manuela, Borsari Leonardo, Aliverti Andrea, Paraskevopoulos Ioannis Th, LoMauro Antonella

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Accurate measurements of limb volumes are important for clinical reasons. We aimed to assess the reliability and validity of two centimetric and two optoelectronic techniques for limb volume measurements against water volumetry, defined as the gold standard.

METHODS

Five different measurement methods were executed on the same day for each participant, namely water displacement, fixed-height (circumferences measured every 5 (10) cm for the upper (lower limb) centimetric technique, segmental centimetric technique (circumferences measured according to proportional height), optoelectronic plethysmography (OEP, based on a motion analysis system), and IGOODI Gate body scanner technology (which creates an accurate 3D avatar).

RESULTS

A population of 22 (15 lower limbs, 11 upper limbs, 8 unilateral upper limb lymphoedema, and 6 unilateral lower limb lymphoedema) participants was selected. Compared to water displacement, the fixed-height centimetric method, the segmental centimetric method, the OEP, and the IGOODI technique resulted in mean errors of 1.2, 0.86, -16.0, and 0.71%, respectively. The corresponding slopes (and regression coefficients) of the linear regression lines were 1.0002 (0.98), 1.0047 (0.99), 0.874 (0.94) and 0.9966 (0.99).

CONCLUSION

The centimetric methods and the IGOODI system are accurate in measuring limb volume with an error of <2%. It is important to evaluate new objective and reliable techniques to improve diagnostic and follow-up possibilities.

Для цитирования: Farina Giovanni, Galli Manuela, Borsari Leonardo, Aliverti Andrea, Paraskevopoulos Ioannis Th, LoMauro Antonella. Limb Volume Measurements: A Comparison of Circumferential Techniques and Optoelectronic Systems against Water Displacement.. Bioengineering (Basel, Switzerland). 2024. DOI: 10.3390/bioengineering11040382.