Successful long-term ventricular pacing via the coronary sinus after the Fontan operation.

Rosenthal E, Qureshi S A, Crick J C

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

A man with double inlet left ventricle and severe subpulmonary stenosis underwent a Fontan operation at the age of 29 years. Eight years later he developed atrial flutter with complete heart block. To avoid a further thoracotomy, a unipolar carbon tipped electrode was placed into the posterior cardiac vein via the coronary sinus. More than 8 years after implantation of the original lead, and after two generator changes, telemetric thresholds remain between 1.8-2.1 volts. Percutaneous transvenous ventricular pacing via the coronary sinus can produce an excellent long-term result and should be the initial approach of choice after a Fontan-type operation.