Measurement Accuracy When Using Spot Vision Screener With or Without Cycloplegia in Young Adults.

Tatara Shunya, Maeda Fumiatsu, Ubukata Hokuto, Shiga Yuko, Yaoeda Kiyoshi

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: There are many unclear points about the accuracy of measurement of cycloplegic refraction using the Spot Vision Screener (SVS). This study aimed to investigate the accuracy of SVS measurements with cycloplegia for myopia.

MATERIALS AND

METHODS

Forty-nine healthy subjects were included, and refraction was measured. Objective refractions were measured by SVS, table-mounted autorefractometer (RT7000), and handheld autorefractometer (Retinomax Screeen) at noncycloplegic and cycloplegic conditions by 1% cyclopentolate. Subjective noncycloplegic refraction was obtained by a visual acuity and refraction test performed by certified orthoptists using a cross-cylinder. One-way repeated-measures analysis of variance was used to examine whether the measured refractions fluctuate due to different reflection tests.

RESULTS

In the noncycloplegic condition, the mean (±standard deviation) spherical equivalent (SE) measured by subjective method, SVS, RT7000, and Retinomax Screeen were -2.56 ± 3.00, -2.62 ± 2.38, -3.05 ± 2.84, and -3.26 ± 2.97, respectively. The subjective SE and objective SE measured by SVS had significantly less myopic value than the objective SE measured by two autorefractometers (p < 0.001). In the cycloplegic condition, the mean (± standard deviation) SE measured by SVS, RT7000, and Retinomax Screeen were -2.07 ± 2.66, -2.62 ± 2.98, and -2.66 ± 3.02, respectively. The objective SE measured by SVS had significantly less myopic value than SEs measured using other methods (p < 0.001). In the cycloplegic condition, SVS showed a fixed error wherein the SE was more hyperopic than that with the subjective method and SVS had a proportional error.

CONCLUSION

In the measurement under cycloplegic conditions, use of an autorefractometer rather than a photorefractometer such as SVS was preferable.

Для цитирования: Tatara Shunya, Maeda Fumiatsu, Ubukata Hokuto, Shiga Yuko, Yaoeda Kiyoshi. Measurement Accuracy When Using Spot Vision Screener With or Without Cycloplegia in Young Adults.. Clinical ophthalmology (Auckland, N.Z.). 2023. DOI: 10.2147/opth.s431202.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.