The Enigmatic Cornea and Intraocular Lens Calculations: The LXXIII Edward Jackson Memorial Lecture.
Аннотация
PURPOSE: To review the progress and challenges in obtaining accurate corneal power measurements for intraocular lens (IOL) calculations.
DESIGN
Personal perspective, review of literature, case presentations, and personal data.
METHODS
Through literature review findings, case presentations, and data from the author's center, the types of corneal measurement errors that can occur in IOL calculation are categorized and described, along with discussion of future options to improve accuracy.
RESULTS
Advances in IOL calculation technology and formulas have greatly increased the accuracy of IOL calculations. Recent reports suggest that over 90% of normal eyes implanted with IOLs may achieve accuracy to within 0.5 diopter (D) of the refractive target. Though errors in estimation of corneal power can cause IOL calculation errors in eyes with normal corneas, greater difficulties in measuring corneal power are encountered in eyes with diseased, scarred, and postsurgical corneas. For these corneas, problematic issues are quantifying anterior corneal power and measuring posterior corneal power and astigmatism. Results in these eyes are improving, but 2 examples illustrate current limitations: (1) spherical accuracy within 0.5 D is achieved in only 70% of eyes with post-refractive surgery corneas, and (2) astigmatism accuracy within 0.5 D is achieved in only 80% of eyes implanted with toric IOLs.
CONCLUSION
Corneal power measurements are a major source of error in IOL calculations. New corneal imaging technology and IOL calculation formulas have improved outcomes and hold the promise of ongoing progress.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.