Astigmatism corrected common path probe for optical coherence tomography.

Singh Kanwarpal, Yamada Daisuke, Tearney Guillermo

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND AND

OBJECTIVES

Optical coherence tomography (OCT) catheters for intraluminal imaging are subject to various artifacts due to reference-sample arm dispersion imbalances and sample arm beam astigmatism. The goal of this work was to develop a probe that minimizes such artifacts.

MATERIALS AND

METHODS

Our probe was fabricated using a single mode fiber at the tip of which a glass spacer and graded index objective lens were spliced to achieve the desired focal distance. The signal was reflected using a curved reflector to correct for astigmatism caused by the thin, protective, transparent sheath that surrounds the optics. The probe design was optimized using Zemax, a commercially available optical design software. Common path interferometric operation was achieved using Fresnel reflection from the tip of the focusing graded index objective lens. The performance of the probe was tested using a custom designed spectrometer-based OCT system.

RESULTS

The probe achieved an axial resolution of 15.6 μm in air, a lateral resolution 33 μm, and a sensitivity of 103 dB. A scattering tissue phantom was imaged to test the performance of the probe for astigmatism correction. Images of the phantom confirmed that this common-path, astigmatism-corrected OCT imaging probe had minimal artifacts in the axial, and lateral dimensions.

CONCLUSIONS

In this work, we developed an astigmatism-corrected, common path probe that minimizes artifacts associated with standard OCT probes. This design may be useful for OCT applications that require high axial and lateral resolutions. Lasers Surg. Med. 49:312-318, 2017. © 2016 Wiley Periodicals, Inc.

Для цитирования: Singh Kanwarpal, Yamada Daisuke, Tearney Guillermo. Astigmatism corrected common path probe for optical coherence tomography.. Lasers in surgery and medicine. 2016. DOI: 10.1002/lsm.22554.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.