DOI 10.1016/j.jcrs.2015.05.042
Measurement agreement between a new biometer based on partial coherence interferometry and a validated biometer based on optical low-coherence reflectometry.
Аннотация
PURPOSE: To evaluate the agreement of ocular measurements obtained with a new optical biometer (AL-Scan) and a previously validated optical biometer (Lenstar).
SETTING
Eye Hospital of Wenzhou Medical University, Wenzhou, Zhejiang, China.
DESIGN
Prospective observational cross-sectional study.
METHODS
For a comprehensive comparison between the partial coherence interferometry (PCI) device and the optical low-coherence reflectometry (OLCR) device, the axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth, mean keratometry (K), astigmatism, white-to-white (WTW), and pupil diameter were measured 3 times per device in eyes with cataract. The sequence of the device was in random order. The mean values were compared and 95% limits of agreement (LoA) were assessed.
RESULTS
Ninety-two eyes of 92 cataract patients were included. Bland-Altman analysis showed excellent agreement between the PCI device and the OLCR device for AL, CCT, ACD and aqueous depth measurements with narrow 95% LoA (-0.05 to 0.06 mm, -13.39 to 15.61 μm, -0.11 to 0.10 mm, and -0.12 to 0.10 mm, respectively), and the P values were more than 0.05. The mean K, astigmatism, and WTW values provided by the PCI device were in good agreement with the OLCR device, although statistically significant differences were detected. A major difference was observed in the pupil diameter measurement, with a 95% LoA of -0.73 to 1.21 mm.
CONCLUSIONS
The PCI device biometer provided ocular measurements similar to those provided by the OLCR device for most parameters, especially for AL, CCT, and ACD. The pupil diameter values obtained with the PCI device were in poor agreement with the OLCR device, and these measurements should be interpreted with necessary adjustment.
FINANCIAL DISCLOSURE: None of the authors has a proprietary or financial interest in any material or method mentioned.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.