Journal of cataract and refractive surgery · 2016

Multicenter study of optical low-coherence interferometry and partial-coherence interferometry optical biometers with patients from the United States and China.

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Hoffer Kenneth J, Shammas H John, Savini Giacomo, Huang Jinhai

Аннотация

PURPOSE: To evaluate the agreement between the measurements provided by a new optical biometer, the Aladdin, based on optical low-coherence interferometry (OLCI), and those provided by the most commonly used optical biometer (IOLMaster 500), based on partial-coherence interferometry (PCI).

SETTING: Multicenter clinical trial.

DESIGN: Prospective evaluation of diagnostic test.

METHODS: In this study, 2 samples of adult patients were enrolled, 1 in the United States and the other in China. The U.S. group included a sample of consecutive patients scheduled for cataract surgery. The China group included a sample of healthy subjects with no cataracts. In both cases, only 1 eye of each patient was analyzed. Axial length (AL), corneal power (in diopters [D]) (K), anterior chamber depth (ACD) (corneal epithelium to lens), and corneal astigmatism were measured. All values were analyzed using a paired t test, the Pearson product-moment correlation coefficient (r), and Bland-Altman plots.

RESULTS: In the U.S. and China groups, the OLCI mean AL values did not show a statistically significant difference from PCI values and showed excellent agreement and correlation. On the contrary, OLCI measured a lower mean K (-0.14 D) and a deeper ACD measurements (U.S. +0.16 mm and China +0.05 mm). These differences were statistically significant (P < .0001). Vector analysis did not show a statistically significant difference in astigmatism measurements.

CONCLUSIONS: Agreement between OLCI and PCI was good. However, the small but statistically significant differences in K and ACD measurements make constant optimization necessary when calculating the intraocular lens power using theoretical formulas.

FINANCIAL DISCLOSURE: Dr. Hoffer licenses the registered trademark name Hoffer to Carl Zeiss-Meditec (PCI), Haag-Streit (Lenstar), Movu (Argos), Oculus (Pentacam, AXL), Nidek (AL-Scan), Tomey (OA-2000), Topcon EU Visia Imaging (Aladdin), Ziemer (Galilei G6), and all A-scan biometer manufacturers. Dr. Shammas licenses his formulas to Carl Zeiss-Meditec (PCI), Haag-Streit (Lenstar), Nidek (AL-Scan), and Topcon EU (Visia Imaging) (Aladdin). None of the other authors has a financial or proprietary interest in any material or method mentioned.