Repeatability of posterior and total corneal curvature measurements with a dual Scheimpflug-Placido tomographer.

Оценочное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Kim Eric J, Montes de Oca Ildamaris, Wang Li, Weikert Mitchell P, Koch Douglas D, Khandelwal Sumitra S

Аннотация

PURPOSE: To evaluate the repeatability of the Galilei G4 dual Scheimpflug analyzer in measuring simulated keratometric, total, and posterior corneal curvature in normal and post-refractive surgery eyes.

SETTING: Cullen Eye Institute, Baylor College of Medicine, Houston, Texas, USA.

DESIGN: Prospective evaluation of diagnostic technology.

METHODS: A single observer performed 3 consecutive measurements in 1 eye of each subject. The following were evaluated in both eyes and in eyes that had previous myopic excimer-laser surgery: (1) simulated keratometric corneal power and astigmatism, (2) total corneal power and astigmatism, and (3) posterior corneal power and astigmatism. Repeatability was assessed by calculating the within-subject standard deviation (Sw), coefficient of variation (CoV), and intra-class correlation coefficient (ICC).

RESULTS: The study evaluated 41 normal eyes and 36 post-refractive surgery eyes. In normal eyes, the Sw was 0.08 diopters (D), 0.10 D, and 0.03 D for simulated keratometric, total, and posterior corneal power, respectively. The CoV ranged from 0.16% to 0.40%, and the ICC was 0.992 or more (P < .001) for all corneal powers. In post-refractive surgery eyes, the Sw was 0.09 D, 0.09 D, and 0.02 D for simulated keratometric, total, and posterior corneal power, respectively. The CoV ranged from 0.19% to 0.32%, and the ICC was 0.990 or more (P < .001) for all corneal powers. For posterior corneal astigmatism, the ICC was 0.814 and 0.886 for normal and post-refractive surgery eyes, respectively.

CONCLUSIONS: In normal corneas and corneas that had undergone myopic excimer laser ablation, the dual Scheimpflug analyzer showed high intra-observer repeatability for simulated keratometric, total, and posterior corneal power measurements and moderate repeatability for posterior corneal astigmatism.

FINANCIAL DISCLOSURE: Drs. Koch, Weikert, and Wang received research support from Ziemer USA, Inc. No other author has a financial or proprietary interest in any material or method mentioned.