Improving perfusion quantification in arterial spin labeling for delayed arrival times by using optimized acquisition schemes.

Kramme Johanna, Gregori Johannes, Diehl Volker, Madai Vince I, von Samson-Himmelstjerna Federico C, Lentschig Markus, Sobesky Jan, Günther Matthias

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

The improvement in Arterial Spin Labeling (ASL) perfusion quantification, especially for delayed bolus arrival times (BAT), with an acquisition redistribution scheme mitigating the T1 decay of the label in multi-TI ASL measurements is investigated. A multi inflow time (TI) 3D-GRASE sequence is presented which adapts the distribution of acquisitions accordingly, by keeping the scan time constant.

MATERIAL AND

METHODS

The MR sequence increases the number of averages at long TIs and decreases their number at short TIs and thus compensating the T1 decay of the label. The improvement of perfusion quantification is evaluated in simulations as well as in-vivo in healthy volunteers and patients with prolonged BATs due to age or steno-occlusive disease.

RESULTS

The improvement in perfusion quantification depends on BAT. At healthy BATs the differences are small, but become larger for longer BATs typically found in certain diseases. The relative error of perfusion is improved up to 30% at BATs>1500ms in comparison to the standard acquisition scheme.

CONCLUSION

This adapted acquisition scheme improves the perfusion measurement in comparison to standard multi-TI ASL implementations. It provides relevant benefit in clinical conditions that cause prolonged BATs and is therefore of high clinical relevance for neuroimaging of steno-occlusive diseases.

Для цитирования: Kramme Johanna, Gregori Johannes, Diehl Volker, Madai Vince I, von Samson-Himmelstjerna Federico C, Lentschig Markus, Sobesky Jan, Günther Matthias. Improving perfusion quantification in arterial spin labeling for delayed arrival times by using optimized acquisition schemes.. Zeitschrift fur medizinische Physik. 2014. DOI: 10.1016/j.zemedi.2014.07.003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.