DOI 10.1097/j.jcrs.0000000000001515
Comparison of keratometry and total corneal power, as measured by an SS-OCT-based optical biometer, for intraocular lens power calculation in Asian eyes.
Аннотация
PURPOSE: To investigate whether standard keratometry (K) or total corneal power (TCP) leads to more accurate refractive outcomes for intraocular lens (IOL) power calculation.
SETTING
Public hospital.
DESIGN
Retrospective evaluation of a diagnostic test instrument.
METHODS
Preoperatively, all patients underwent optical biometry with Anterion, a swept-source optical coherence tomography device providing both K and TCP. The same IOL model was implanted in all cases. The whole sample was divided into a training dataset, used to optimize the formula constants, and a testing dataset, used to investigate the spherical equivalent prediction error (SEQ-PE) of 8 IOL power formulas. Trueness, precision, and accuracy were evaluated by means of the robust 2-sample t test. Cochran Q test was performed to assess whether the percentage of eyes with an SEQ-PE within each threshold was significantly different; in such an event, the McNemar test was then applied.
RESULTS
Both the training and testing datasets included 317 eyes. No significant differences were detected for trueness because of constant optimization. Precision and accuracy were better when K was entered, although a statistically significant difference was observed only with the EVO (precision: P = .02 and accuracy: P = .03) and Haigis ( P < .01 for both precision and accuracy) formulas. No significant differences were observed for the percentage of eyes with an absolute SEQ-PE within any threshold.
CONCLUSIONS
With most formulas, IOL power calculation is not more accurate when TCP is used instead of K.