Comparison of keratometry and total corneal power, as measured by an SS-OCT-based optical biometer, for intraocular lens power calculation in Asian eyes.

Savini Giacomo, Taroni Leonardo, Jeon Sohee, Koh Kyungmin, Yang Hyun Seung, Abulafia Adi, Lupardi Enrico, Hoffer Kenneth J, Moramarco Antonio, Schiano-Lomoriello Domenico

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To investigate whether standard keratometry (K) or total corneal power (TCP) leads to more accurate refractive outcomes for intraocular lens (IOL) power calculation.

SETTING

Public hospital.

DESIGN

Retrospective evaluation of a diagnostic test instrument.

METHODS

Preoperatively, all patients underwent optical biometry with Anterion, a swept-source optical coherence tomography device providing both K and TCP. The same IOL model was implanted in all cases. The whole sample was divided into a training dataset, used to optimize the formula constants, and a testing dataset, used to investigate the spherical equivalent prediction error (SEQ-PE) of 8 IOL power formulas. Trueness, precision, and accuracy were evaluated by means of the robust 2-sample t test. Cochran Q test was performed to assess whether the percentage of eyes with an SEQ-PE within each threshold was significantly different; in such an event, the McNemar test was then applied.

RESULTS

Both the training and testing datasets included 317 eyes. No significant differences were detected for trueness because of constant optimization. Precision and accuracy were better when K was entered, although a statistically significant difference was observed only with the EVO (precision: P = .02 and accuracy: P = .03) and Haigis ( P < .01 for both precision and accuracy) formulas. No significant differences were observed for the percentage of eyes with an absolute SEQ-PE within any threshold.

CONCLUSIONS

With most formulas, IOL power calculation is not more accurate when TCP is used instead of K.

Для цитирования: Savini Giacomo, Taroni Leonardo, Jeon Sohee, Koh Kyungmin, Yang Hyun Seung, Abulafia Adi, Lupardi Enrico, Hoffer Kenneth J, Moramarco Antonio, Schiano-Lomoriello Domenico. Comparison of keratometry and total corneal power, as measured by an SS-OCT-based optical biometer, for intraocular lens power calculation in Asian eyes.. Journal of cataract and refractive surgery. 2024. DOI: 10.1097/j.jcrs.0000000000001515.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.