Intraocular pressure before and after capsulorhexis using two viscoelastic substances and two surgical approaches in enucleated porcine eyes.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ben Abdallah Chiraz, Sekundo Walter, Ladewig Markus S, Handzel Daniel M
Аннотация
AIM: To investigate the influence of ophthalmic viscoelastic devices (OVDs) and different surgical approaches on the intraocular pressure (IOP) before and after creation of the curvilinear circular capsulorhexis (CCC) as a measure for anterior chamber stability during this maneuver.
METHODS: Prospective experimental WetLab study carried out on enucleated porcine eyes. IOP was measured before and after CCC with the iCare Rebound tonometer (iCare ic200; iCare Finland Oy, Vantaa, Finland). The OVDs used were a cohesive one [Z-Hyalin, Carl Zeiss Meditec AG, Germany; hyaluronic acid (HA)] and a dispersive [Z-Celcoat, Carl Zeiss Meditec AG, Germany; hydroxy propylmethylcellulosis (HPMC)]. The CCC was created using Utrata forceps or 23 g microforceps in different combinations with the OVDs.
RESULTS: Using the Utrata forceps the IOP dropped from 63.65±6.44 to 11.25±3.63 mm Hg during the CCC. The use of different OVDs made no difference. Using the 23 g microforceps the IOP dropped from 65.35±8.15 to 36.55±6.09 mm Hg. The difference between IOP drop using either Utrata forceps or 23 g microforceps was highly significant regardless of the OVD used.
CONCLUSION: Using the sideport for the creation of the capsulorhexis leads to a lesser drop in IOP during this maneuver compared to the main incision in enucleated porcine eyes. The use of different OVD has no significant influence on IOP drop.
METHODS: Prospective experimental WetLab study carried out on enucleated porcine eyes. IOP was measured before and after CCC with the iCare Rebound tonometer (iCare ic200; iCare Finland Oy, Vantaa, Finland). The OVDs used were a cohesive one [Z-Hyalin, Carl Zeiss Meditec AG, Germany; hyaluronic acid (HA)] and a dispersive [Z-Celcoat, Carl Zeiss Meditec AG, Germany; hydroxy propylmethylcellulosis (HPMC)]. The CCC was created using Utrata forceps or 23 g microforceps in different combinations with the OVDs.
RESULTS: Using the Utrata forceps the IOP dropped from 63.65±6.44 to 11.25±3.63 mm Hg during the CCC. The use of different OVDs made no difference. Using the 23 g microforceps the IOP dropped from 65.35±8.15 to 36.55±6.09 mm Hg. The difference between IOP drop using either Utrata forceps or 23 g microforceps was highly significant regardless of the OVD used.
CONCLUSION: Using the sideport for the creation of the capsulorhexis leads to a lesser drop in IOP during this maneuver compared to the main incision in enucleated porcine eyes. The use of different OVD has no significant influence on IOP drop.