DOI 10.12968/jowc.2023.32.4.247
Cold plasma treatment is safe for diabetic foot ulcers and decreases Staphylococcus aureus bacterial load.
Аннотация
AIM: Cold atmospheric plasma (CAP) has antimicrobial properties. We studied the safety of a novel CAP device (PLASOMA prototype; Plasmacure, The Netherlands) that is simple to use and could be applied at a patient's home for the treatment of diabetic foot ulcers (DFUs). Secondary objectives were to investigate the effect of CAP on bacterial load and on ulcer size.
METHOD: We included subjects with non-infected, superficial DFUs and treated them with CAP on a daily basis for 10 days. The primary endpoint was the occurrence of serious adverse device effects (SADE). We defined safety as: ≤10% of patients experiencing a SADE other than infection (non-infectious SADE), and ≤60% of patients developing infection of the foot (infectious serious adverse event (SAE)).
RESULTS
We enrolled 20 patients. No SADE occurred, but three infectious SAEs occurred at the site of application within one month of treatment; three SAEs unrelated to treatment occurred, and 55% of subjects reported transient mild adverse device effects. Staphylococcus aureus bacterial load decreased directly after CAP application (p=0.01). The mean decrease of ulcer surface area was 43% (95% confidence interval: 20.2%-65.9%).
CONCLUSION
CAP treatment in DFUs was safe and well tolerated. Ulcer size and Staphylococcus aureus colonisation decreased during treatment.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.