Clinically important factors influencing the diagnostic measurement of pleural fluid pH and glucose.
Аннотация
RATIONALE: Accurate pleural fluid pH and glucose measurement is a key component in the diagnosis and management of patients with pleural effusion. Standardized methods of pleural fluid collection have not been defined.
OBJECTIVES
To assess the effect of common clinical factors that may distort measurement accuracy of pleural fluid pH and glucose.
METHODS
Ninety-two exudative pleural aspirates were collected in commercially available blood gas syringes.
MEASUREMENTS AND MAIN
RESULTS
Samples were analyzed immediately using a blood gas analyzer. The effects of residual air, lidocaine, heparin, and delay in analysis (24 h) on pH and glucose measurement accuracy were assessed. Pleural fluid pH was significantly increased by residual air (mean +/- SD, 0.08 +/- 0.07; 95% confidence interval [CI], 0.06 to 0.09; P < 0.001) and significantly decreased by residual lidocaine (0.2 ml; mean change in pH, -0.15 +/- 0.09; 95% CI, -0.13 to -0.18; P < 0.001) and residual heparin (mean change in pH, -0.02 +/- 0.05; 95% CI, -0.01 to -0.04; P = 0.027). Pleural fluid pH was stable at room temperature for 1 hour and significantly increased at 4 (mean +/- SD, 0.03 +/- 0.07; 95% CI, 0.01 to 0.04; P = 0.003) and 24 hours (0.05 +/- 0.12; 95% CI, 0.03 to 0.08; P < 0.001). Pleural fluid glucose concentration was not clinically significantly altered by residual air, lidocaine (up to 0.4 ml), or 24-hour analysis delay.
CONCLUSIONS
Accuracy of measured pleural pH is critically dependent on sample collection method. Residual air, lidocaine, and analysis delay significantly alter pH and may impact on clinical management. Pleural fluid glucose concentration is not significantly influenced by these factors. Protocols defining appropriate sampling and analysis methods are needed.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.