MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/j.1540-8159.1986.tb06654.x

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 1986

Electrode configurations for a respiratory impedance measurement suitable for rate responsive pacing.

Nappholtz T, Valenta H, Maloney J, Simmons T

СтатьяПеревод пока не подготовлен

Аннотация

A rate responsive cardiac pacemaker will benefit the patient only if the input parameter is a reliable indicator of exercise. Part of the reliability is dependent upon selection of the best physiological parameter and part is dependent upon the measurement technique. Twenty-six patients have been studied in order to ascertain the best measurement configuration for an intravascular impedance respiratory monitor. Four electrode configurations have been examined at rest and during exercise. The results have been compared using measures of sensitivity and specificity. The ideal configuration requires that a current field be established between the right ventricular blood and the pacemaker case. And, that voltage sensing be done between an electrode in the SVC and the pacemaker case.

Для цитирования: Nappholtz T, Valenta H, Maloney J, Simmons T. Electrode configurations for a respiratory impedance measurement suitable for rate responsive pacing.. Pacing and clinical electrophysiology : PACE. 1986. DOI: 10.1111/j.1540-8159.1986.tb06654.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.