m+Medelevator.RU
← Поиск публикаций
Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2003

The use of full-thickness skin grafts for the skin-abutment interface around bone-anchored hearing aids.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Snyder Mary C, Moore Gary F, Johnson Perry J

Аннотация

OBJECTIVE: To review the complication rate encountered with the use of full-thickness skin grafts to establish the skin-abutment interface around bone-anchored hearing aid implants.

STUDY DESIGN: Retrospective chart review.

SETTING: Tertiary referral center.

PATIENTS: Fifteen patients who underwent bone-anchored hearing aid placement over a 4-year period.

INTERVENTION: Each percutaneous titanium implant and abutment was placed into the temporal bone following the standard Branemark technique. Eight procedures were performed in two stages, and seven were performed as single-stage procedures. In all cases, the skin-abutment interface was established by use of a full-thickness skin graft inset around the implant.

MAIN OUTCOME MEASURES: The incidence of complications resulting in skin graft loss, time from implantation to bone-anchored hearing aid use, additional procedures for revision of the interface, and complicating medical factors in the patient population.

RESULTS: Seven patients (46.7%) experienced loss of the full-thickness skin graft around the abutment. Four of these seven had complicating medical factors associated with impaired wound healing: two with diabetes mellitus, one of whom was also a smoker, and two patients who were receiving inhaled steroids for treatment of asthma. Of the seven patients who lost skin grafts, two healed by secondary intention, two underwent repeated full-thickness skin grafting, and three underwent galeal rotation flaps with split-thickness skin grafting, one of which eventually required a scalp flap. No patient experienced loss of the implant.

CONCLUSION: The use of full-thickness skin grafts for establishment of the skin-abutment interface around bone-anchored hearing aid implants is associated with a high rate of graft loss. Although salvage techniques can successfully establish the interface after skin graft failure, alternative methods should be considered, especially in high-risk patients.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты