Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2001

Automatic atrial anti-tachy pacing for the termination of spontaneous atrial tachyarrhythmias: clinical experience with a novel dual-chamber pacemaker.

Клиническое исследованиеСравнительное исследованиеОценочное исследованиеСтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Vollmann D, Stevens J, Buchwald A B, Unterberg C

Аннотация

UNLABELLED: Automatic atrial anti-tachy pacing (aATP) is a novel approach to treat paroxysmal/persistent atrial tachyarrhythmias in pacemaker patients. To evaluate the efficacy of aATP in terminating spontaneous atrial flutter/tachycardia episodes (AT), a dual-chamber stimulator with extensive diagnostic capabilities and programmable aATP therapies (AT500(TM), Medtronic Inc.) was implanted in 30 patients with conventional pacing indications. During a mean follow-up time of 5.5 (1-12) months, aATP was delivered for 10494 AT. According to automatic device analysis, 8289 AT were treated with success (success-rate 79.0%). On 468 AT stored with the corresponding atrial EGM, an additional manual analysis was performed. The success-rate based on automatic analysis of these AT episodes (73.1%) was comparable to that found for all treated AT (79.0%), but manual EGM analysis revealed that only 209 of the 468 treated AT episodes (44.7%) were actually terminated by aATP. The aATP success-rate in the slower (cycle length 360-270 ms) AT detection zone was significantly higher (73.8%, 62/84 eps) than in the overlapping, faster (cycle length 270-220 ms) AT zone (38.3%, 147/384 eps, P<0.01).

CONCLUSIONS: According to manual analysis, 1. aATP was safe and had a success-rate of 44.7%, 2. aATP success-rate was higher for AT in the slower than in the faster detection zone and 3. automatic analysis overestimated the efficacy of aATP.