Lessons from individualized cryoballoon sizing. Is there a role for the small balloon?

Hartl Stefan, Dorwarth Uwe, Bunz Benedikt, Wankerl Michael, Ebersberger Ullrich, Hoffmann Ellen, Straube Florian

Клиническое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Cryoablation for paroxysmal atrial fibrillation (PAF) is well established. The single-big-balloon strategy has been preferred for pulmonary vein isolation (PVI) using the second generation cryoballoon (CBG2). Individual PV-morphologies raise the question if an individualized anatomic approach using the 23-mm or 28-mm CB is reasonable.

METHODS

Consecutive patients were prospectively enrolled in the non-randomized single-center study. Patients were treated with the 28-mm CB, if any PV was >21mm, the 23-mm CB, if all PV were ≤21mm, or both sizes, if PVI was difficult. The primary endpoint was arrhythmia-free survival. The secondary endpoint considered procedural results and complications.

RESULTS

Overall, 197 patients with symptomatic PAF (64±11 years, 36% female) were included. Acute PVI was achieved in 99.9% of PV. Based on preprocedural imaging, the 28-mm CB was applied as the primary catheter in 47% (92/197 patients), the 23-mm CB in 53% (105/197, p=0.23). The 23-mm CB group included more females, patients with short left atrial (LA)-diameters (each p<0.01), and smaller patients (p=0.04). Both CB-sizes were used in 24% (47/197). Additional 23-mm CB usage was necessary in 23% (21/92) of patients, mainly because of insufficient PV-occlusion with the 28-mm CB. Additional 28-mm CB usage was necessary in 25% (26/105, p=0.82), mainly because PV diameters were larger than initially measured. Both CB-sizes were equally safe and effective with a low complication rate and an overall success rate of 86% at 12 and 71% at 18 months (6% on antiarrhythmic drugs). No predictors for AF-recurrence were identified.

CONCLUSION

CB ablation can sometimes be challenging. The 28-mm CB is the preferred catheter in all patients. If balloon positioning is difficult, the 23-mm CB is an option to achieve PVI in small veins. Further studies need to investigate if the 23-mm CB could be beneficial as the primary CB in females with small body height and short LA diameter.

Для цитирования: Hartl Stefan, Dorwarth Uwe, Bunz Benedikt, Wankerl Michael, Ebersberger Ullrich, Hoffmann Ellen, Straube Florian. Lessons from individualized cryoballoon sizing. Is there a role for the small balloon?. Journal of cardiology. 2017. DOI: 10.1016/j.jjcc.2016.12.016.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.