Influence of energy source on early atrial fibrillation recurrences: a comparison of cryoballoon vs. radiofrequency current energy ablation with the endpoint of unexcitability in pulmonary vein isolation.

Gunawardene Melanie A, Hoffmann Boris A, Schaeffer Benjamin, Chung Da-Un, Moser Julia, Akbulak Ruken Oezge, Jularic Mario, Eickholt Christian, Nuehrich Jana, Meyer Christian, Willems Stephan

Сравнительное исследованиеСтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

INTRODUCTION

Comparative data of early recurrence rates of atrial fibrillation (ERAF) following second-generation cryoballoon (CB-G2) and radiofrequency current (RFC) ablation for pulmonary vein isolation (PVI) in paroxysmal AF (PAF) are rare. We randomized PAF patients into either PVI with CB-G2 (group 1) or PVI with a combined RFC-approach applying contact force (CF) with the endpoint of unexcitability (group 2) to investigate ERAF.

METHODS AND

RESULTS

In group 1 (n = 30), CB-G2-PVI was performed. After CF-PVI in group 2 (n = 30), bipolar pacing on the ablation line and additional ablation until unexcitability was conducted. Follow-up included 48 h of in-hospital monitoring followed by 5-day Holter ECGs 1, 2, 3, 6, 12 months postablation to evaluate ERAF. Acute PVI was reached in 100% of group 2 and in 99% of group 1. Shorter procedure durations (98.0 ± 21.9 vs. 114.3 ± 18.7 min, P < 0.05) but extended fluoroscopy times (15.4 ± 3.9 vs. 10.0 ± 4.3 min, P < 0.05) were found in the CB-G2 group. Ten non-severe complications occurred (6 vs. 4 in group 1 and 2, P = 0.73). In group 2, five patients suffered from ERAF vs. seven patients in group 1 (P = 0.67). The time until the occurrence of ERAF was shorter in group 2 (1 day (q1-q3: 1-4.5)) when compared with group 1 (22 (q1-q3: 6-54) days, P = 0.025).

CONCLUSION

ERAF rates were equal among groups; however, they occurred earlier in the initial phase after RFC ablation when compared with CB-G2. PVI utilizing cryoablation is associated with shorter procedure durations but extended fluoroscopy time while being similarly secure.

Для цитирования: Gunawardene Melanie A, Hoffmann Boris A, Schaeffer Benjamin, Chung Da-Un, Moser Julia, Akbulak Ruken Oezge, Jularic Mario, Eickholt Christian, Nuehrich Jana, Meyer Christian, Willems Stephan. Influence of energy source on early atrial fibrillation recurrences: a comparison of cryoballoon vs. radiofrequency current energy ablation with the endpoint of unexcitability in pulmonary vein isolation.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2018. DOI: 10.1093/europace/euw307.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.