Fiberoptic bronchoscopy during noninvasive positive pressure ventilation delivered by helmet.
Аннотация
OBJECTIVE
To evaluate the feasibility and safety of fiberoptic bronchoscopy (FOB) with bronchoalveolar lavage (BAL) during noninvasive positive pressure ventilation (NPPV) delivered by helmet in patients with acute respiratory failure (ARF) and suspected pneumonia.
DESIGN AND
SETTING
Prospective, clinical investigation in a general intensive care unit (ICU) of a university hospital.
PATIENTS AND
PARTICIPANTS
Four adult patients with ARF who underwent NPPV via the helmet and required fiberoptic BAL for suspected pneumonia.
INTERVENTIONS: NPPV was delivered through the helmet in the pressure support ventilation mode. The specific seal connector placed in the plastic ring of the helmet allowed the passage of the bronchoscope, maintaining assisted ventilation. Arterial blood gas levels, pH, oxygen saturation, respiratory rate, heart rate, and mean arterial blood pressure were monitored during the study.
RESULTS
Helmet NPPV avoided gas exchanges deterioration during FOB and BAL, with good tolerance. During the procedure heart rate increased by 5% and mean arterial blood pressure by 7% over baseline; these levels returned to prebronchoscopic values immediately after the withdrawal of the bronchoscope. Endotracheal intubation was never required during the 24 h after the procedure. BAL yielded diagnostic information in three of four patients.
CONCLUSIONS
NPPV through the helmet allows a safe diagnostic FOB with BAL in patients with hypoxemic ARF, avoiding gas exchange deterioration, and endotracheal intubation.
ELECTRONIC SUPPLEMENTARY MATERIAL: is available if you access this article at http://dx.doi.org/10.1007/s00134-002-1554-5. On that page (frame on the left side), a link takes you directly to the supplementary materials.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.