Comparison of the effects of heat and moisture exchangers and heated humidifiers on ventilation and gas exchange during non-invasive ventilation.

Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Jaber Samir, Chanques Gérald, Matecki Stefan, Ramonatxo Michèle, Souche Bruno, Perrigault Pierre-François, Eledjam Jean-Jacques

Аннотация

OBJECTIVE: To compare the short-term effects of a heat and moisture exchanger (HME) and a heated humidifier (HH) during non-invasive ventilation (NIV).

DESIGN: Prospective, clinical investigation.

SETTING: Intensive care unit of a university hospital.

PATIENTS: Twenty-four patients with acute respiratory failure (ARF).

INTERVENTION: Each patient was studied with a HME and a HH in a random order during two consecutive 20min periods of NIV.

MEASUREMENTS AND RESULTS: Respiratory rate (RR), expiratory tidal volume (VTe) and expiratory minute ventilation (VE) were measured during the last 5 min of each period and blood gases were measured. Mean pressure support and positive end-expiratory pressure levels were, respectively, 15+/-4 and 6+/-2 cmH(2)O. VE was significantly greater with HME than with HH (14.8+/-4.8 vs 13.2+/-4.3 l/min; p<0.001). This increase in VE was the result of a greater RR for HME than for HH (26.5+/-10.6 vs 24.1+/-9.8 breaths/min; p=0.002), whereas the VT for HME was similar to that for HH (674+/-156 vs 643+/-148 ml; p=0.09). Arterial partial pressure of carbon dioxide (PaCO(2)) was significantly higher with a HME than with a HH (43.4+/-8.9 vs 40.8+/-8.2 mmHg; p<0.005), without significantly changing oxygenation.

CONCLUSION: During NIV the increased dead space of a HME can negatively affect ventilatory function and gas exchange. The effect of HME dead space may decrease efficiency of NIV in patients with ARF.