MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ijid.2021.11.022

Медицинская библиотека · International journal of infectious diseases : IJID : official publication of the International Society for Infectious Diseases · 2021

Nationwide external quality assessment of SARS-CoV-2 nucleic acid amplification tests in Japan.

Asai Satomi, Seki Akira, Akai Yasumasa, Tazawa Hiromitsu, Kakizoe Hidehumi, Ravzanaaadii Mend-Amar, Miyachi Hayato

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

We conducted a nationwide external quality assessment (EQA) study of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) nucleic acid amplification testing in Japan.

METHODS

A total of 563 public health and private sector laboratories participated. The EQA samples comprised 6 RNA and full-process controls.

RESULTS

The overall agreements were 99.3% and 97.9% for the RNA and full-process controls, respectively. A total of 530/563 (94.1%) laboratories reported correct results; public health laboratories had the highest accuracy. Thirty-three laboratories reported at least one incorrect result (26 laboratories of medical facilities, 5 commercial laboratories, 1 public health laboratory, and 1 other). Sixteen laboratories of medical facilities that used a fully automated assay system failed to detect the presence of the full-process control, due to inherent insufficiency in the limit of detection (LOD). Other causes of incorrect results included failure to ensure the LOD (n = 13), error in result judging or reporting (n = 3), and error in sample handling (n = 1).

CONCLUSIONS

Performance was mostly dependent on the laboratory category and assay evaluation, particularly the LOD. Guidance should be developed based on these results, particularly in the phase of new entry into laboratory services for SARS-CoV-2.

Для цитирования: Asai Satomi, Seki Akira, Akai Yasumasa, Tazawa Hiromitsu, Kakizoe Hidehumi, Ravzanaaadii Mend-Amar, Miyachi Hayato. Nationwide external quality assessment of SARS-CoV-2 nucleic acid amplification tests in Japan.. International journal of infectious diseases : IJID : official publication of the International Society for Infectious Diseases. 2021. DOI: 10.1016/j.ijid.2021.11.022.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.