DOI 10.3109/02713683.2014.941071
The effect of collagen cross-linking procedure on the material of intracorneal ring segments.
Аннотация
PURPOSE: To assess the potential impact of corneal crosslinking treatment (365 nm ultraviolet (UV)A irradiation with riboflavin) on the material properties of Intracorneal Rings Segments (ICRS).
MATERIALS AND
METHODS
Material properties were studied using FT-IR spectroscopy and UV-Vis spectroscopy. Rings were examined: (1) after installation of riboflavin solution, (2) after irradiation with UV-A light and (3) after instillation of riboflavin solution followed by instillation and coeval irradiation. The experiments followed the standard corneal cross-linking (CXL) protocol of corneal crosslinking treatment.
RESULTS
After instillation of riboflavin solution, a permanent intense yellow staining of the samples was observed. UV-Vis spectroscopy confirmed that a certain amount of riboflavin solution was absorbed into the samples after CXL procedure. FT-IR spectroscopic analysis showed alterations in the spectra of ICRS mainly at the 2800-3200 cm(-1) spectral region [modification in band intensities of CH(2) (2925 cm(-1)) and CH(3) (2950 cm(-1))].
CONCLUSIONS
Our results suggest crosslinking reaction in ICRS material. This should be taken into consideration prior to any CXL treatment of post ICRS-implanted cornea.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.