Optical coherence tomography combined with videokeratography to differentiate mild keratoconus subtypes.

Rabinowitz Yaron S, Li Xiaohui, Canedo Ana Laura Caiado, Ambrósio Renato, Bykhovskaya Yelena

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To develop parameters using a combination of optical coherence tomography (OCT) and videokeratography to detect early keratoconus.

METHODS

Videokeratography, wavefront analysis, and measured OCT indices were performed on 180 normal eyes, 46 eyes with moderate keratoconus, 54 eyes with early keratoconus, 7 eyes with forme fruste keratoconus, and 16 eyes with keratoconus "suspect" to determine the most sensitive parameters for separating these groups.

RESULTS

A combination of videokeratography and OCT indices (inferior-superior [I-S] value and minimum pachymetry) was statistically the most significant in separating the keratoconus groups from normal eyes (P < .001). Using a newly derived index, the minimum pachymetry divided by the I-S value (pachymetry/asymmetry [PA]/I-S index) with a cut-off of 100, 100% of early and forme fruste keratoconus could be identified as being abnormal with 7 normals misclassified (misclassification rate 2.7%). By adding keratoconus "suspect" to the analysis and an I-S value of 1.2 as a cut-off point, 5 "suspects" were classified as normal and 11 normals as abnormal (misclassification rate 7.8%). The PA/I-S index, with a cut-off point of 100, reduced this misclassification rate to 4.4%.

CONCLUSIONS

These results suggest that OCT combined with videokeratography may be more useful for differentiating mild forms of keratoconus than videokeratography alone.

Для цитирования: Rabinowitz Yaron S, Li Xiaohui, Canedo Ana Laura Caiado, Ambrósio Renato, Bykhovskaya Yelena. Optical coherence tomography combined with videokeratography to differentiate mild keratoconus subtypes.. Journal of refractive surgery (Thorofare, N.J. : 1995). 2014. DOI: 10.3928/1081597x-20140120-02.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.