Corneal stroma demarcation line after standard and high-intensity collagen crosslinking determined with anterior segment optical coherence tomography.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Kymionis George D, Tsoulnaras Konstantinos I, Grentzelos Michael A, Plaka Argyro D, Mikropoulos Dimitrios G, Liakopoulos Dimitrios A, Tsakalis Nikolaos G, Pallikaris Ioannis G
Аннотация
PURPOSE: To use anterior segment optical coherence tomography (AS-OCT) to compare corneal stroma demarcation line depth after corneal collagen crosslinking (CXL) with 2 treatment protocols.
SETTING: Vardinoyiannion Eye Institute of Crete, Faculty of Medicine, University of Crete, Heraklion, Crete, Greece.
DESIGN: Prospective comparative interventional case series.
METHODS: Corneal collagen crosslinking was performed in all eyes using the same ultraviolet-A (UVA) irradiation device (CCL-365). Eyes were treated for 30 minutes with 3 mW/cm(2) according to the standard Dresden protocol (Group 1) or for 10 minutes with 9 mW/cm(2) of UVA irradiation intensity (Group 2). One month postoperatively, 2 independent observers measured the corneal stroma demarcation line using AS-OCT.
RESULTS: Sixteen patients (21 eyes) were enrolled. Group 1 comprised 7 patients (9 eyes) and Group 2, 9 patients (12 eyes). The mean corneal stroma demarcation line depth was 350.78 μm ± 49.34 (SD) (range 256.5 to 410 μm) in Group 1 and 288.46 ± 42.37 μm (range 238.5 to 353.5 μm) in Group 2; the corneal stroma demarcation line was statistically significantly deeper in Group 1 than in Group 2 (P=.0058, t test for unpaired data).
CONCLUSION: The corneal stroma demarcation line was significantly deeper after a 30-minute CXL treatment than after a 10-minute CXL procedure with high-intensity UVA irradiation.
SETTING: Vardinoyiannion Eye Institute of Crete, Faculty of Medicine, University of Crete, Heraklion, Crete, Greece.
DESIGN: Prospective comparative interventional case series.
METHODS: Corneal collagen crosslinking was performed in all eyes using the same ultraviolet-A (UVA) irradiation device (CCL-365). Eyes were treated for 30 minutes with 3 mW/cm(2) according to the standard Dresden protocol (Group 1) or for 10 minutes with 9 mW/cm(2) of UVA irradiation intensity (Group 2). One month postoperatively, 2 independent observers measured the corneal stroma demarcation line using AS-OCT.
RESULTS: Sixteen patients (21 eyes) were enrolled. Group 1 comprised 7 patients (9 eyes) and Group 2, 9 patients (12 eyes). The mean corneal stroma demarcation line depth was 350.78 μm ± 49.34 (SD) (range 256.5 to 410 μm) in Group 1 and 288.46 ± 42.37 μm (range 238.5 to 353.5 μm) in Group 2; the corneal stroma demarcation line was statistically significantly deeper in Group 1 than in Group 2 (P=.0058, t test for unpaired data).
CONCLUSION: The corneal stroma demarcation line was significantly deeper after a 30-minute CXL treatment than after a 10-minute CXL procedure with high-intensity UVA irradiation.