m+Medelevator.RU
← Поиск публикаций
International journal of ophthalmology · 2013

Measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Yazıcı Ahmet Taylan, Pekel Gökhan, Bozkurt Ercüment, Yıldırım Yusuf, Pekel Evre, Demirok Ahmet, Yılmaz Omer Faruk

Аннотация

AIM: To evaluate the measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients.

METHODS: A hundred and one eyes of 55 keratoconus patients were enrolled in this study. The mean age was 26.2±8.9 years. The inclusion criteria were keratoconus stage I to III according to the Amsler-Krumeich keratoconus classification. All the measurements were done by the same operator, under the mesopic light condition and repeated with three different optical methods; Visante, Orbscan and Pentacam. The evaluated anterior segment parameters were anterior chamber depth (ACD), central and thinnest corneal thickness (CCT and TCT) and pupil diameter (PD).

RESULTS: THE MEAN CCT MEASURED BY VISANTE, ORBSCAN AND PENTACAM WERE AS FOLLOWS: 462.0±48.1µm, 463.9±60.9µm, 476.5±45.3µm, respectively (P=0.873). The mean ACD values were 3.34±0.33mm, 3.26±0.33mm, 3.49±0.40mm, respectively (P=0.118). The mean PD measurements were 5.11±1.14mm, 4.80±0.85mm, 3.80±1.38mm, respectively (P<0.001). The mean TCT measurements of Visante, Orbscan and Pentacam were 437.9±48.2µm, 447.6±60.6µm and 459.9±44.0µm, respectively (P=0.214). The Visante and Orbscan measured CCT similarly, while Pentacam measured CCT thicker than the other two. The Visante measured TCT thinner than the other two devices. In ACD measurements, Orbscan was the one giving the lowest values. PD was measured differently by the devices.

CONCLUSION: Although TCT, CCT and ACD measurements acquired by Visante, Orbscan and Pentacam in keratoconus patients are similar, PD measurements show large differences among the devices.

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты