Efficacy of corneal curvature on the accuracy of 8 intraocular lens power calculation formulas in 302 highly myopic eyes.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Mo Er, Feng Ke, Li Qiyuan, Xu Jingyuan, Cen Jiaying, Li Jin, Zhao Yun-E
Аннотация
PURPOSE: To investigate the effect of corneal curvature (K) on the accuracy of 8 intraocular lens formulas in highly myopic eyes.
SETTING: Eye Hospital and School of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, China.
DESIGN: Retrospective consecutive case series.
METHODS: 302 eyes (302 patients) were analyzed in subgroups based on the K value. The mean refractive error, mean absolute error (MAE), median absolute error (MedAE), root-mean-square absolute prediction error (RMSAE) and proportions of eyes within ±0.25 diopter (D), ±0.50 D, ±0.75 D, ±1.00 D were statistical analyzed.
RESULTS: Emmetropia Verifying Optical (EVO) 2.0, Kane, and Radial Basis Function (RBF) 3.0 had the lower MAE (≤0.28) and RMSAE (≤0.348) and highest percentage of eyes within ±0.50 D (≥83.58%) in the flat (K ≤ 43 D) and steep K (K > 45 D) groups. Hoffer QST had the lowest MedAE (0.19), RMSAE (0.351) and the highest percentage of eyes within ±0.50 D (82.98%) in the normal K group (43 < K ≤ 45 D). When axial length (AL) ≤28 mm, all formulas showed close RMSAE values (0.322 to 0.373) in flat K group. When AL >28 mm, RBF 3.0 achieved the lowest MAE (≤0.24), MedAE (≤0.17) and RMSAE (≤0.337) across all subgroups.
CONCLUSIONS: EVO 2.0, Kane, and RBF 3.0 were the most accurate in highly myopic eyes with a flat or steep K. Hoffer QST is recommended for long eyes with normal K values. RBF 3.0 showed the highest accuracy when AL >28 mm, independent of corneal curvature.
SETTING: Eye Hospital and School of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, China.
DESIGN: Retrospective consecutive case series.
METHODS: 302 eyes (302 patients) were analyzed in subgroups based on the K value. The mean refractive error, mean absolute error (MAE), median absolute error (MedAE), root-mean-square absolute prediction error (RMSAE) and proportions of eyes within ±0.25 diopter (D), ±0.50 D, ±0.75 D, ±1.00 D were statistical analyzed.
RESULTS: Emmetropia Verifying Optical (EVO) 2.0, Kane, and Radial Basis Function (RBF) 3.0 had the lower MAE (≤0.28) and RMSAE (≤0.348) and highest percentage of eyes within ±0.50 D (≥83.58%) in the flat (K ≤ 43 D) and steep K (K > 45 D) groups. Hoffer QST had the lowest MedAE (0.19), RMSAE (0.351) and the highest percentage of eyes within ±0.50 D (82.98%) in the normal K group (43 < K ≤ 45 D). When axial length (AL) ≤28 mm, all formulas showed close RMSAE values (0.322 to 0.373) in flat K group. When AL >28 mm, RBF 3.0 achieved the lowest MAE (≤0.24), MedAE (≤0.17) and RMSAE (≤0.337) across all subgroups.
CONCLUSIONS: EVO 2.0, Kane, and RBF 3.0 were the most accurate in highly myopic eyes with a flat or steep K. Hoffer QST is recommended for long eyes with normal K values. RBF 3.0 showed the highest accuracy when AL >28 mm, independent of corneal curvature.