DOI 10.1097/ftd.0000000000000632

Медицинская библиотека · Therapeutic drug monitoring · 2019

Interlaboratory Analysis of Isavuconazole Plasma Concentration Assays Among European Laboratories.

Pea Federico, Krause Robert, Müller Carsten, Hennart Benjamin, Richardson Malcolm, Meinitzer Andreas, Wiesen Martin H J, Wiktorowicz Tatiana, Spickermann Jochen, Henriksen Anne S

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Under certain circumstances, clinicians treating patients with isavuconazole for invasive aspergillosis or mucormycosis may use therapeutic drug monitoring. However, the accuracy and reproducibility of the various assays used by different laboratories for the quantification of isavuconazole plasma concentrations have yet to be determined.

METHODS

Human plasma samples spiked with known concentrations of isavuconazole were provided to 27 European laboratories that took part in a "round-robin" test (an interlaboratory test performed independently at least 2 times; 2 rounds performed in the current study). Assay methods included liquid chromatography-tandem mass spectrometry (LC-MS/MS), LC with ultraviolet detection (LC-UV), LC with fluorescence detection (LC-FL), and bioassay. The accuracy and reproducibility compared with the known concentrations for each sample in each round were compared overall, between assays, and between laboratories.

RESULTS

Twenty-seven laboratories participated in the study (LC-MS/MS, n = 15; LC-UV; n = 9; LC-FL, n = 1; bioassay, n = 2). In round 1, for nominal concentrations of 1000, 1700, 2500, and 4000 ng/mL, the mean (SD) determined concentrations were 1007 (183), 1710 (323), 2528 (540), and 3898 (842) ng/mL, respectively. In round 2, for nominal concentrations of 1200, 1800, 2400, and 4000 ng/mL, the mean (SD) determined concentrations were 1411 (303), 2111 (409), 2789 (511), and 4723 (798) ng/mL, respectively. Over both rounds, determined concentrations were consistently within 15% of the nominal concentrations for 10 laboratories (LC-MS/MS, n = 4; LC-UV, n = 5; bioassay, n = 1) and consistently exceeded the upper 15% margin for 7 laboratories (LC-MS/MS and LC-UV, n = 3 each; LC-FL, n = 1).

CONCLUSIONS

Alignment of methodologies among laboratories may be warranted to improve the accuracy and reproducibility of therapeutic drug measurements.

Для цитирования: Pea Federico, Krause Robert, Müller Carsten, Hennart Benjamin, Richardson Malcolm, Meinitzer Andreas, Wiesen Martin H J, Wiktorowicz Tatiana, Spickermann Jochen, Henriksen Anne S. Interlaboratory Analysis of Isavuconazole Plasma Concentration Assays Among European Laboratories.. Therapeutic drug monitoring. 2019. DOI: 10.1097/ftd.0000000000000632.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.