Fluoroscopic angiography quantifies delay in cerebral circulation time and requires less radiation in carotid stenosis patients: A pilot study.

Chu Wei-Fa, Lee Han-Jui, Lin Chung-Jung, Chang Feng-Chi, Guo Wan-Yuo, Chen Liang-Wei, Lin Yi-Yang, Luo Chao-Bao

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Quantitative digital subtraction angiography (DSA) facilitates in-room assessment of flow changes in various cerebrovascular diseases and improves patient safety. The purpose of this study was to compare the diagnostic accuracy of quantitative fluoroscopic angiography (FA) and DSA.

METHODS

Twenty-two patients with >70% carotid stenosis according to NASCET criteria were prospectively included in the study. All patients received DSA and FA (ArtisZee, Siemens Healthcare, Forchheim, Germany) before and after carotid stenting in the same angiosuite. The regions of interest (ROIs) included the extracranial internal carotid artery (eICA), first segment of the middle cerebral artery (MCA1), and sigmoid sinus in the anterior-posterior view; cavernous portion of the ICA (cICA), parietal vein, and jugular vein in the lateral views. The time-to-peak (TTP) for all ROIs and cerebral circulation time (CCT) were measured from FA and DSA scans. TTP, CCT, and radiation doses from DSA were compared with those from FA.

RESULTS

The mean age of the patients were 69 ± 9.5 years old. The average stenosis was 89.7% ± 7.8% before stenting and 31% ± 3.6% after stenting. No patient suffered from periprocedural stroke. The intermethod correlation for TTP for all ROIs except the eICA and cICA ranged from 0.46 to 0.65 before stenting and 0.57 to 0.73 after stenting, and that for CCT was 0.65 before stenting and 0.57 after stenting. The radiation doses were significantly lower for FA than for DSA regardless of views or periprocedural timing (p < 0.001).

CONCLUSION

Stenosis facilitated the creation of a bolus by manual injection and therefore increased the accuracy of cerebral flow quantification in FA. Cerebral hemodynamic assessment by FA is quicker and associated with less radiation.

Для цитирования: Chu Wei-Fa, Lee Han-Jui, Lin Chung-Jung, Chang Feng-Chi, Guo Wan-Yuo, Chen Liang-Wei, Lin Yi-Yang, Luo Chao-Bao. Fluoroscopic angiography quantifies delay in cerebral circulation time and requires less radiation in carotid stenosis patients: A pilot study.. Journal of the Chinese Medical Association : JCMA. 2019. DOI: 10.1097/jcma.0000000000000046.