MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00261-015-0608-4

Медицинская библиотека · Abdominal radiology (New York) · 2016

Optimization of non-contrast-enhanced MR angiography of the renal artery with three-dimensional balanced steady-state free-precession and time-spatial labeling inversion pulse (time-SLIP) at 3T MRI, in relation to age and blood velocity.

Kurata Yasuhisa, Kido Aki, Fujimoto Koji, Kiguchi Kayo, Takakura Kyoko, Moribata Yusaku, Shitano Fuki, Himoto Yuki, Fushimi Yasutaka, Okada Tomohisa, Togashi Kaori

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To determine the optimal inversion time (TI) value of three-dimensional (3D) balanced steady-state free-precession time-spatial labeling inversion pulse (time-SLIP) technique for visualization of the renal artery at 3T MRI, and to assess whether the optimal TI is affected by the subject's age and blood velocity.

MATERIALS AND

METHODS

Forty-two healthy volunteers (range 20-67 years) were enrolled in the study and subjected to non-contrast-enhanced renal MR angiography. Five different TI values (1200, 1400, 1600, 1800, and 2000 ms) were selected for evaluation. For quantitative evaluation, the relative signal intensity (SI) of the main renal artery was compared with that of the renal medulla (Vessel-to-Kidney ratio; VKR). Blood velocity of the abdominal aorta was measured using 2D phase contrast technique. For qualitative evaluation, two radiologists scored the depiction of the renal pelvis and the quality of visualization of the renal artery.

RESULTS

VKR is the highest at TI = 1600 ms. A strong negative correlation between age and blood velocity was demonstrated. Regarding the qualitative evaluation, the overall image scores of renal arteries were the highest at a TI = 1800 ms for both readers. The optimal TI values in subjects below 50 years of age were 1600 and 1800 ms, whereas in subjects above 50 years of age, the optimal TI value was 1800 ms.

CONCLUSION

The optimal TI value for the visualization of renal arteries using time-SLIP technique at 3T MRI was 1800 ms. Subjects' age affected optimal TI and this is likely due to differences in the blood velocity of the abdominal aorta.

Для цитирования: Kurata Yasuhisa, Kido Aki, Fujimoto Koji, Kiguchi Kayo, Takakura Kyoko, Moribata Yusaku, Shitano Fuki, Himoto Yuki, Fushimi Yasutaka, Okada Tomohisa, Togashi Kaori. Optimization of non-contrast-enhanced MR angiography of the renal artery with three-dimensional balanced steady-state free-precession and time-spatial labeling inversion pulse (time-SLIP) at 3T MRI, in relation to age and blood velocity.. Abdominal radiology (New York). 2016. DOI: 10.1007/s00261-015-0608-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.