DOI 10.1016/j.jcrs.2014.12.060

Медицинская библиотека · Journal of cataract and refractive surgery · 2015

Correction of low levels of astigmatism.

Bullimore Mark A, Spooner Greg, Sluyterman Georg, Dishler Jon G

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To reevaluate the analysis of the correction of astigmatism.

SETTING

Academia, industry, and private practice.

DESIGN

Evaluation of diagnostic test or technology.

METHODS

Astigmatic refractive surgery outcomes are based on vector methods, including the correction index (also known as the correction ratio), which is the ratio of the surgically induced astigmatism to the target induced astigmatism (TIA). Mean correction indices substantially greater than 1 have been reported for astigmatic corrections less than 1.00 diopter (D) and as representing systematic overcorrection. We hypothesize that this reflects a limitation of the correction index rather than systematic flaws in treatments. The theoretical mathematic behavior of the correction index was analyzed, accounting for variability in astigmatism measurement. Then, the impact of cylinder measurement variability on the mean correction index was modeled. A Monte Carlo simulation was performed and calculated 10 000 values of correction index for various values of TIA. Finally, correction indices from published and unpublished studies of refractive lasers were compared with the simulations.

RESULTS

The mean correction index is always greater than 1 for the case of a perfect refractive correction; however, for astigmatic corrections less than 1.00 D, the mean correction index increases sharply because the measurement variability is similar in magnitude to TIA. Almost all previous studies show the predicted increase in the correction index for low astigmatic corrections.

CONCLUSION

The correction index is a useful vector-based metric for the evaluation of refractive procedures, but mean values greater than 1 should be anticipated for lower astigmatic treatments and do not necessarily represent systematic overcorrection.

FINANCIAL DISCLOSURE: Dr. Bullimore is a consultant to Alcon Surgical, Inc., Carl Zeiss Meditec AG, Digital Vision Systems, Essilor, Innovega, Inc., and Paragon Vision Sciences, Inc. Dr. Spooner is a consultant to Alcon Surgical, Inc., Carl Zeiss Meditec AG, Digital Vision Systems, Thru-Focus Optics LLC, and i2eyediagnostics, Ltd. Dr. Dishler is a consultant to Carl Zeiss Meditec AG and Revision Optics, Inc. Dr. Sluyterman is an employee of Carl Zeiss Meditec AG.

Для цитирования: Bullimore Mark A, Spooner Greg, Sluyterman Georg, Dishler Jon G. Correction of low levels of astigmatism.. Journal of cataract and refractive surgery. 2015. DOI: 10.1016/j.jcrs.2014.12.060.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.