Retrospective Analysis Comparing the Preoperative and Postoperative "Q" Values for 2 Different Lasers in Refractive Surgery.

Molchan Ryan P, Taylor Kenneth R, Panday Vasudha A, Caldwell Matthew C, Reilly Charles D

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the presurgical and postsurgical corneal asphericity, or "Q value," between VISX Wavefront-guided treatment and Allegretto Wave Wavefront-optimized treatment in photorefractive keratectomy for the correction of myopia.

METHODS

This is a retrospective database analysis performed at the Joint Warfighter Refractive Surgery Center, Lackland Air Force Base, TX. All data were accessed through the Institutional Review Board-approved Joint Warfighter Refractive Surgery Center database. Fifty-one patient records (102 eyes) were reviewed to determine the change in the Q value after photorefractive keratectomy for the correction of myopia.

RESULTS

The average change in the Q value per diopter of treatment sphere over the entire study population was 0.12 (±0.04) for Allegretto and 0.14 (±0.04) for VISX (P = 0.004). There was no statistical difference among the preoperative and postoperative best-corrected visual acuities (BCVAs), low-contrast BCVA (BCVA 5%), or quality of vision complaints between both platforms.

CONCLUSIONS

Low to moderate myopic correction with the Allegretto laser was associated with a smaller change in the Q value per diopter of treatment and therefore more closely maintained the original prolate shape of the cornea when compared with VISX. No difference in the postoperative visual outcome between both groups was detected.

Для цитирования: Molchan Ryan P, Taylor Kenneth R, Panday Vasudha A, Caldwell Matthew C, Reilly Charles D. Retrospective Analysis Comparing the Preoperative and Postoperative "Q" Values for 2 Different Lasers in Refractive Surgery.. Cornea. 2015. DOI: 10.1097/ico.0000000000000611.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.