Intraobserver repeatability and interobserver reproducibility of corneal measurements in normal eyes using an optical coherence tomography-Placido disk device.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Shah Jinesh M, Han Daphne, Htoon Hla Myint, Mehta Jodhbir S
Аннотация
PURPOSE: To study intraobserver and interobserver reproducibility of corneal topography and aberrometry measurements using a device combining optical coherence tomography (OCT) and a Placido disk (Visante Omni).
SETTING: Singapore National Eye Centre, Singapore.
DESIGN: Prospective nonrandomized study.
METHOD: Measurements of simulated keratometry (K), anterior mean curvature, spherical anterior and posterior elevations, 4.5 mm diameter zone aberrometry (spherical aberration, asphericity, astigmatism), pachymetry, and white-to-white (WTW) diameter were taken in normal right eyes. Mean biases and limits of agreement (LoA) were obtained using Bland-Altman analysis.
RESULTS: Regarding intraobserver repeatability, the mean biases for simulated K, anterior mean curvature, spherical anterior/posterior elevation, 4.5 mm zone aberrometry, pachymetry, and WTW diameter were not statistically significant (P > .05). The LoA for the anterior mean curvature measurements were flat K, -3.96 diopters (D) and 4.86 D; steep K, -4.50 D and 4.27 D; mean K, -2.48 D and 2.09 D. The LoA for WTW were -9.90 mm and 8.74 mm. Regarding interobserver reproducibility, the mean biases for simulated K, anterior mean curvature, spherical anterior/posterior elevation, 4.5 mm zone aberrometry, pachymetry, and WTW diameter were not statistically significant (P > .05). The LoA for anterior mean curvature were flat K, -4.05 D and 4.13 D; steep K, -5.08 D and 3.85 D; mean K, -2.65 D and 2.29 D. The LoA for WTW diameter were -8.79 mm and 8.24 mm.
CONCLUSIONS: The OCT-Placido device had high intraobserver and interobserver clinical reproducibility for measurements of simulated K, 4.5 mm zone aberrometry, pachymetry, and standard anterior/posterior corneal elevation. However, intraobserver and interobserver reproducibility of anterior mean curvature measurements and WTW diameter was lower.
FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.
SETTING: Singapore National Eye Centre, Singapore.
DESIGN: Prospective nonrandomized study.
METHOD: Measurements of simulated keratometry (K), anterior mean curvature, spherical anterior and posterior elevations, 4.5 mm diameter zone aberrometry (spherical aberration, asphericity, astigmatism), pachymetry, and white-to-white (WTW) diameter were taken in normal right eyes. Mean biases and limits of agreement (LoA) were obtained using Bland-Altman analysis.
RESULTS: Regarding intraobserver repeatability, the mean biases for simulated K, anterior mean curvature, spherical anterior/posterior elevation, 4.5 mm zone aberrometry, pachymetry, and WTW diameter were not statistically significant (P > .05). The LoA for the anterior mean curvature measurements were flat K, -3.96 diopters (D) and 4.86 D; steep K, -4.50 D and 4.27 D; mean K, -2.48 D and 2.09 D. The LoA for WTW were -9.90 mm and 8.74 mm. Regarding interobserver reproducibility, the mean biases for simulated K, anterior mean curvature, spherical anterior/posterior elevation, 4.5 mm zone aberrometry, pachymetry, and WTW diameter were not statistically significant (P > .05). The LoA for anterior mean curvature were flat K, -4.05 D and 4.13 D; steep K, -5.08 D and 3.85 D; mean K, -2.65 D and 2.29 D. The LoA for WTW diameter were -8.79 mm and 8.24 mm.
CONCLUSIONS: The OCT-Placido device had high intraobserver and interobserver clinical reproducibility for measurements of simulated K, 4.5 mm zone aberrometry, pachymetry, and standard anterior/posterior corneal elevation. However, intraobserver and interobserver reproducibility of anterior mean curvature measurements and WTW diameter was lower.