Clinical results with a supplementary toric intraocular lens for the correction of astigmatism in pseudophakic patients.
Аннотация
PURPOSE: To evaluate the visual outcomes of pseudophakic patients who underwent supplementary toric intraocular lens (IOL) implantation to correct astigmatic refractive errors.
METHODS
Pseudophakic patients referred for the implantation of a supplementary toric IOL (Sulcoflex Toric 653T) were evaluated. Uncorrected and corrected distance visual acuities (UDVA and CDVA, respectively), spherical equivalent (SE) refraction, rotational stability, higher order aberrations (HOA), and photopic glare and no-glare contrast sensitivity (CSV-1000, VectorVision) were evaluated.
RESULTS
A total of 10 eyes of 10 patients were included. The mean age was 56.42 ± 5.9 years (range 45-65 years). Mean follow-up was 6.99 ± 5.1 months (6-18 months). Postoperatively, UDVA improved to 0.10 ± 0.12 (0.3 to -0.1) (p = 0.004) and CDVA to 0.07 ± 0.12 (0.3 to -0.1) (p = 0.021). Mean SE was -0.30 ± 0.56 D (-1.00 to +0.75) (p = 0.001). Mean toric IOL axis rotation at 6-month follow-up was 3.0° ± 2.45° (0-6). Ocular aberrometry values decreased after surgery (for average HOA root mean square, p = 0.008). Photopic contrast sensitivity (for all spatial frequencies) showed a trend for improvement after surgery; however, this was not borne out from the analysis (p>0.05).
CONCLUSIONS
The implantation of the Sulcoflex Toric IOL to correct astigmatism in pseudophakic patients provided excellent visual outcomes, predictability of refractive results, rotational stability, and optical performance. The implantation of this IOL is a safe and effective technique to correct pseudophakic cylindrical refractive errors and reduce spectacle dependence in these patients.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.