Surgical Efficiency Comparison Between Two Phacoemulsification Systems.
Аннотация
PURPOSE: To assess the surgical efficiency in cataract surgery comparing two phacoemulsification systems.
METHODS
Prospective, consecutive-comparative study in a two-site private practice. Three hundred and one eyes undergoing standard or femtosecond laser-assisted (FLACS) cataract surgery with either the R-Evo Smart (BVI, Waltham, USA) and/or the Centurion Vision System (Alcon, Fort Worth, USA). Preoperative eye characteristics (degree of cataract using the lens opacities classification system LOCS III grading) and intraoperative outcomes (total ultrasound time and total estimated fluid aspirated/drainage bag weighting) were registered in all cases.
RESULTS
One hundred and fifty-five eyes undergone cataract surgery with the R-Evo Smart and 146 eyes with the Centurion Vision System phacoemulsification systems. Mean cataract grade was 3.07 ± 0.78 and 2.96 ± 0.85 for the R-Evo Smart and Centurion Vision System groups, respectively (p = 0.12). Mean total ultrasound time was 18.99 ± 12.85 and 40.24 ± 21.91 seconds for the R-Evo Smart and Centurion Vision System groups, respectively (p < 0.01). Mean total estimated fluid aspirated/drainage bag weighting was 53.00 ± 14.56 g and 54.33 ± 14.88 cc for the R-Evo Smart and Centurion Vision System groups, respectively (p = 0.21). Considering non-FLACS surgery (98 eyes with the R-Evo Smart and 63 eyes with the Centurion Vision System), mean cataract grade was 2.95 ± 0.74 and 2.97 ± 0.91 for the R-Evo Smart and Centurion Vision System groups, respectively (p = 0.44). Mean total ultrasound time was 19.96 ± 11.20 and 42.84 ± 28.35 seconds for the R-Evo Smart and Centurion Vision System groups, respectively (p < 0.01). Mean total estimated fluid aspirated/drainage bag weighting was 55.95 ± 14.76 g and 55.97 ± 13.62 cc for the R-Evo Smart and Centurion Vision System groups, respectively (p = 0.49). No adverse events were found in the two groups of eyes.
CONCLUSION
The objective measurement of surgical efficiency through total ultrasound time during lens removal and fluid consumption during both lens removal and irrigation/aspiration proved R-Evo Smart to be an efficient phacoemulsification platform, in comparison with the current standard of care Centurion Vision System.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.