PURPOSE: Evaluate the impact of high vs low intraoperative intraocular pressure (IOP) during phacoemulsification on anterior and posterior physiology.

SETTING

Private practice in Des Moines, Iowa.

DESIGN

Prospective, randomized, paired-eye clinical trial of patients anticipating bilateral cataract surgery.

METHODS

27 participants randomized at the time of their first cataract surgery to either low IOP or high IOP group. The subsequent cataract surgery was performed under the alternate condition. During phacoemulsification and cortex removal, IOP was maintained either at low (≤28 mm Hg) or at high (≥55 to 60 mm Hg) levels. The primary outcome was fluid usage, with secondary outcomes of central corneal thickness, foveal avascular zone area, foveal and macular thickness, endothelial cell density (ECD), and postoperative inflammation.

RESULTS

Surgery with low IOP settings used less fluid (40.0 mL vs 55.6 mL, P < .0001). Corneal thickness changes were smaller in low IOP eyes at 1 day and 1 week (3.0% vs 8.1%, P = .01; 3.1% vs 4.4%, P = .01) but were similar by 1 and 3 months. ECD dropped less in low IOP eyes at 1 and 3 months (-1.7% vs -12.3%, P = .001, 2.1% vs -8.9%, P = .0003.) IOP remained a significant predictor of ECD change when relationship was controlled for fluid use and phacoemulsification energy. Retinal parameters did not vary among all eyes or when compared by IOP setting. Visual acuity was similar at all timepoints.

CONCLUSIONS

Low IOP settings resulted in less inflammation and less corneal trauma, as evidenced by a smaller drop in ECD and less postoperative corneal edema, when compared with high IOP settings. Retinal parameters did not change significantly. The different outcomes did not result in a difference in visual acuity.

Для цитирования: Rauen Matthew P, Joiner Hillery, Kohler Robyn A, O'Connor Sara. Phacoemulsification using an active fluidics system at physiologic vs high intraocular pressure: impact on anterior and posterior segment physiology.. Journal of cataract and refractive surgery. 2024. DOI: 10.1097/j.jcrs.0000000000001457.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.