Potential role of the light-adjustable lens in flanged intrascleral haptic fixation.

LoBue Stephen A, Saba Nicholas, Catapano Thomas M, Martin Curtis R, Shelby Christopher L, Coleman Wyche T

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To compare maximum tensile strength between commonly used 3-piece intraocular lens (IOL) for flanged intrascleral haptic fixation (FISHF).

SETTING

Willis-Knight Eye Institute, Shreveport, Louisiana.

DESIGN

Laboratory investigation.

METHODS

Haptic tensile strength was compared with MA60AC, CT Lucia 602, AR40E, and the light-adjustable lens (LAL). Haptic strength with a 24-diopter (D) IOL was compared across all lenses, as well as across a range of 10 to 30 D with the MA60AC. A custom device was created to hold the IOL in correct haptic orientation. The maximum tension (mean ± SD) was recorded in Newtons (N) when the haptic lost tension or broke.

RESULTS

CT Lucia was the strongest at 1.53 ± 0.11 N vs 1.00 ± 0.15 (MA60AC), 0.87 ± 0.19 (AR40E), and 0.83 ± 0.14 N (LAL) ( P < .001). The LAL and AR40E were similar to a 9-0 polypropylene suture while being significantly stronger than 10-0 polypropylene suture ( P < .001). No difference in haptic tension for the MA60AC from 10 to 30 D ( P > .05). High magnification revealed the highest haptic fractures for MA60AC at 40% compared with LAL, AR40E, and CT Lucia at 0%. CT Lucia and AR40E had 100% of haptics disinserted from the IOL without any damage compared with 60% LAL and 60% MA60AC. CT Lucia, AR40E, and LAL have a flatter haptic angulation at 5 degrees.

CONCLUSIONS

Haptic strength, durability, and angulation of the LAL may support the possibility of FISHF in the hands of experienced surgeons. However, further testing is strongly recommended to verify whether physiologic conditions or light treatments may compromise long-term haptic stability.

Для цитирования: LoBue Stephen A, Saba Nicholas, Catapano Thomas M, Martin Curtis R, Shelby Christopher L, Coleman Wyche T. Potential role of the light-adjustable lens in flanged intrascleral haptic fixation.. Journal of cataract and refractive surgery. 2024. DOI: 10.1097/j.jcrs.0000000000001455.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.