A multi-sensor wearable system for the assessment of diseased gait in real-world conditions.

Salis Francesca, Bertuletti Stefano, Bonci Tecla, Caruso Marco, Scott Kirsty, Alcock Lisa, Buckley Ellen, Gazit Eran, Hansen Clint, Schwickert Lars, Aminian Kamiar, Becker Clemens, Brown Philip, Carsin Anne-Elie, Caulfield Brian, Chiari Lorenzo, D'Ascanio Ilaria, Del Din Silvia, Eskofier Bjoern M, Garcia-Aymerich Judith, Hausdorff Jeffrey M, Hume Emily C, Kirk Cameron, Kluge Felix, Koch Sarah, Kuederle Arne, Maetzler Walter, Micó-Amigo Encarna M, Mueller Arne, Neatrour Isabel, Paraschiv-Ionescu Anisoara, Palmerini Luca, Yarnall Alison J, Rochester Lynn, Sharrack Basil, Singleton David, Vereijken Beatrix, Vogiatzis Ioannis, Della Croce Ugo, Mazzà Claudia, Cereatti Andrea, For The Mobilise-D Consortium

СтатьяПеревод пока не подготовлен

Аннотация

Introduction

Accurately assessing people's gait, especially in real-world conditions and in case of impaired mobility, is still a challenge due to intrinsic and extrinsic factors resulting in gait complexity. To improve the estimation of gait-related digital mobility outcomes (DMOs) in real-world scenarios, this study presents a wearable multi-sensor system (INDIP), integrating complementary sensing approaches (two plantar pressure insoles, three inertial units and two distance sensors).

Methods

The INDIP technical validity was assessed against stereophotogrammetry during a laboratory experimental protocol comprising structured tests (including continuous curvilinear and rectilinear walking and steps) and a simulation of daily-life activities (including intermittent gait and short walking bouts). To evaluate its performance on various gait patterns, data were collected on 128 participants from seven cohorts: healthy young and older adults, patients with Parkinson's disease, multiple sclerosis, chronic obstructive pulmonary disease, congestive heart failure, and proximal femur fracture. Moreover, INDIP usability was evaluated by recording 2.5-h of real-world unsupervised activity. Results and discussion: Excellent absolute agreement (ICC >0.95) and very limited mean absolute errors were observed for all cohorts and digital mobility outcomes (cadence ≤0.61 steps/min, stride length ≤0.02 m, walking speed ≤0.02 m/s) in the structured tests. Larger, but limited, errors were observed during the daily-life simulation (cadence 2.72-4.87 steps/min, stride length 0.04-0.06 m, walking speed 0.03-0.05 m/s). Neither major technical nor usability issues were declared during the 2.5-h acquisitions. Therefore, the INDIP system can be considered a valid and feasible solution to collect reference data for analyzing gait in real-world conditions.

Для цитирования: Salis Francesca, Bertuletti Stefano, Bonci Tecla, Caruso Marco, Scott Kirsty, Alcock Lisa, Buckley Ellen, Gazit Eran, Hansen Clint, Schwickert Lars, Aminian Kamiar, Becker Clemens, Brown Philip, Carsin Anne-Elie, Caulfield Brian, Chiari Lorenzo, D'Ascanio Ilaria, Del Din Silvia, Eskofier Bjoern M, Garcia-Aymerich Judith, Hausdorff Jeffrey M, Hume Emily C, Kirk Cameron, Kluge Felix, Koch Sarah, Kuederle Arne, Maetzler Walter, Micó-Amigo Encarna M, Mueller Arne, Neatrour Isabel, Paraschiv-Ionescu Anisoara, Palmerini Luca, Yarnall Alison J, Rochester Lynn, Sharrack Basil, Singleton David, Vereijken Beatrix, Vogiatzis Ioannis, Della Croce Ugo, Mazzà Claudia, Cereatti Andrea, For The Mobilise-D Consortium. A multi-sensor wearable system for the assessment of diseased gait in real-world conditions.. Frontiers in bioengineering and biotechnology. 2023. DOI: 10.3389/fbioe.2023.1143248.