DOI 10.1167/tvst.14.3.6

Медицинская библиотека · Translational vision science & technology · 2025

Advances in OCT Angiography.

Hormel Tristan T, Huang David, Jia Yali

СтатьяОбзорПеревод пока не подготовлен

Аннотация

Optical coherence tomography angiography (OCTA) is a signal processing and scan acquisition approach that enables OCT devices to clearly identify vascular tissue down to the capillary scale. As originally proposed, OCTA included several important limitations, including small fields of view relative to allied imaging modalities and the presence of confounding artifacts. New approaches, including both hardware and software, are solving these problems and can now produce high-quality angiograms from tissue throughout the retina and choroid. Image analysis tools have also improved, enabling OCTA data to be quantified at high precision and used to diagnose disease using deep learning models. This review highlights these advances and trends in OCTA technology, focusing on work produced since 2020.

Другие публикации по теме

Tomography, Optical CoherenceAngiographyRetinal VesselsImage Processing, Computer-Assisted
Для цитирования: Hormel Tristan T, Huang David, Jia Yali. Advances in OCT Angiography.. Translational vision science & technology. 2025. DOI: 10.1167/tvst.14.3.6.

Перевод полного текста: Перевод в очереди

Полный текст источника

https://creativecommons.org/licenses/by/4.0/ This work is licensed under a Creative Commons Attribution 4.0 International License.

Текст сохранён. Часть иллюстраций пока доступна только в первоисточнике; загрузка будет проверена повторно.

Introduction

From its introduction, optical coherence tomography (OCT) 1 had clear potential as an ophthalmic imaging modality. OCT offers volumetric imaging and can observe details similar to histology, including the identification of several different retinal tissue layers. 2 However, unlike histology, it can be applied noninvasively to the living eye. Consequently, it is now routinely used for the detection of treatment indicators, including exudation in age-related macular degeneration 3 and diabetic macular edema, 4 and to monitor treatment response. 5

Because tissue contrast in OCT is determined by reflectivity, one key limitation of the approach, as originally proposed, is difficulty visualizing capillaries and small vessels. Microvasculature has similar, though not identical, reflectance to surrounding static tissue, which means that the reflectance signal, and therefore standard structural OCT, is not ideal for identifying microvascular structure, although techniques based on artificial intelligence can help with this task. 6 This is unfortunate because the vasculature that supplies the retina is uniquely intricate 7 and a key component of vision processing. 8 Also, many prevalent vision-threatening diseases include vascular pathology. 9

OCT angiography (OCTA) is a means of supplementing structural OCT with vascular imaging. It works by trading reflectance contrast for motion contrast since blood flow can differentiate static tissue from vessels. In OCTA imaging, we refer to the motion contrast signal mainly attributable to blood flow as the flow signal, and its measurement allows the identification of vascular structure in the retina down to the capillary scale. However, because other sources of motion (e.g., ocular pulsation) also manifest as motion contrast, distinguishing the flow signal from artifacts is a major issue in OCTA. The first OCT demonstrations relied on time-domain processing. 1 , 10 However, scanning rates in these systems yield imaging timescales over which sources of motion (for example, bulk motion and cardiac pulsation) extraneous to blood flow can overwhelm the flow signal. The Fourier domain approach to OCT 11 – 14 eventually produced systems more than a hundred times faster than contemporary time-domain approaches, an improvement that allowed the flow signal to be isolated from other confounding sources of motion. Another important advancement was the realization that calculating motion contrast cross-section by cross-section (B-scan by B-scan) 15 , 16 , rather than line-scan by line-scan (A-scan by A-scan), 17 produced angiograms with higher flow detection sensitivity. One more notable development was the introduction of highly efficient flow signal generation algorithms. 18 , 19 With these approaches, procedure times could be reduced to manageable lengths, and OCTA was ready for the clinic.

The first commercial OCTA devices offered fields of view confined to 3 × 3-mm field of view at high scan density, quantification consisted mostly of vessel density measurements, and images had to be manually graded. Nowadays, we are used to much larger fields of view (usually achieved by lowering scan density by some degree), several different quantifications can be accurately performed from OCTA data, and artificial intelligence algorithms can detect features with high precision. This review will describe and explain advances in OCTA, focusing on work since 2020 to demonstrate how the technology has grown from its introduction and elucidate its contemporary capabilities.

Advances in OCTA Acquisition

The increase in scan acquisition rate provided by Fourier domain relative to time domain processing made OCTA possible, and scanning speed enhancement continues to be a driving force in OCTA technology. With faster data acquisition, developers can choose among several means to improve OCTA images. These include enlarging the field of view, improving sampling resolution, and increasing the dynamic range of flow detection. These improvements are interrelated so that once a field of view and interscan time interval are specified, sampling density is also largely determined.

Most contemporary research devices and many commercial systems use swept-source lasers (SS-OCT). 14 , 20 , 21 These systems can typically achieve higher scan rates than alternative spectral domain (SD-OCT) systems 12 , 22 – 24 ; experimental SS-OCT systems are capable of obtaining multi-MHz scan rates 25 – 27 and the fastest commercial system clocks in at 400-kHz. To the best of our knowledge, the current record holder for the fastest SS-OCTA system uses a stretch-pulse active mode-locked laser and obtains a 9.4-MHz line rate, but this system was only demonstrated on rodent brain images and not the retina. 26 SS-OCTA systems also typically use longer wavelength light sources, which benefits tissue penetration. The result is that SS-OCTA can enlarge both the transverse field of view and imaging depth for an overall larger imaging volume; however, it worth noting that the price of these systems remains relatively high due to the swept source laser component.

Wide-Field OCTA

Restricted fields of view were an important limitation in the first OCTA devices. With a given scan rate, instrument design must balance the triple bind of procedure time, field of view, and sampling density. Early wide-field OCTA studies used montage, usually including several OCTA scans at different locations. 28 – 34 This approach is okay for research but cannot be easily adapted to clinical practice.

Improved scanning speeds from SS-OCTA offer a way to extend the field of view without impractical procedure times. For example, an SS-OCTA system equipped with a 1.68-MHz Fourier-domain mode-locked laser can achieve a 23 × 17-mm field of view with Montage. 35 Jia's group developed a prototype 400-kHz SS-OCTA system (short cavity swept-source laser; Axsun Technologies, Billerica, MA, USA), which demonstrated a single-shot 23 × 12-mm field of view 31 ( Fig. 1 ) in retinal clinical studies. It has been recently upgraded to a 500-kHz SS-OCTA system (vertical cavity surface-emitting laser; Thorlabs, Newton, NJ, USA) with improved sensitivity, generating wide-field OCTA in both superficial and deep slabs. 34 , 36 Single-shot systems can reduce procedure complexity and are also useful for patients who may have difficulty with fixation. These properties are useful for clinical practice, and commercial systems can now also perform wide-field single-shot imaging (15 × 15-mm for Zeiss PLEX Elite [Zeiss, Oberkochen, Germany], 26 × 21-mm for Intalight DREAM [Intalight, San Jose, CA, USA]) in something like two minutes including acquisition and processing (depending on the patients ability to fixate and other external factors), but they achieve this by downsampling.

Иллюстрация — в первоисточнике ↗

Figure 1. A single-shot OCTA en face image of an eye with DR across a 23 × 12 mm field of view, using a wide-field OCTA system developed by Wei et al. 31 This system incorporates a self-tracking method to remove disrupting motion artifacts, 32 and OCTA processing was performed using phase-stabilized complex decorrelation, an approach that efficiently achieves a high flow detection sensitivity. 33 Pathologic features include non-perfusion area ( teal ), retinal neovascularization ( magenta vessels ), and microaneurysms ( red, with dotted circles ). Most of the pathology in this image is located outside of a conventional central macular field of view, demonstrating the usefulness of wide-field imaging in this disease. Reproduced with permission from Hormel et al. 34

Capturing high-quality wide-field images is not as simple as just incorporating a high-speed swept source laser into a device. One issue is that as fields of view increase distortions due to mapping a curved surface to a flat display (like a computer screen) become more severe; these can lead to significant differences in feature quantification and should be corrected. 37 A second is ensuring that the device sample arm covers a sufficiently large range to image the eye at different depths, which also becomes more difficult in widefield imaging. Another problem exacerbated by wide-field imaging is unwanted eye motion (for example due to ocular pulsation, microsaccade, drift, or blinking) and is a particularly acute issue for OCTA because the technique is based on using motion contrast to identify vessels. Because wider fields of view often require longer procedures times motion due to sources other than blood flow can become more severe. Many motion correction approaches rely on allied imaging modalities such as scanning laser ophthalmoscopy and others, examples being Vienola et al. 38 – 40 These approaches are examples of active tracking and complicate imaging systems by requiring additional hardware. As an alternative, contemporary systems can rely on real-time OCTA processing for motion correction 41 or use passive tracking in which problematic regions are re-scanned by the device. 32 These approaches are aided by the use of efficient and sensitive OCTA processing algorithms, for example phase-stabilized complex decorrelation OCT angiography. 33 Registering across orthogonal scanning (as, for example, in Optovue devices [Optovue, Fremont, CA, USA]) directions is another way to address motion artifacts volumetrically ( Fig. 2 ). 42 , 43

Иллюстрация — в первоисточнике ↗

Figure 2. Volumetric motion correction using orthogonal scanning. Left to right: X-fast, Y-fast, merged OCT, and merged OCTA volumes at equal transverse and axial resolution. With permission from Ploner et al. 43

There are obvious reasons for seeking to enlarge the field of view in OCTA imaging. To begin with, important pathology can be located in the periphery. Examples are lesions in retinopathy of prematurity 44 and retinal neovascularization (RNV) in DR. 36 Wide-field OCTA is an advantageous technology for RNV because it can identify neovascularization that is not noticeable in clinical examination, 45 – 47 and in head-to-head comparisons, wide-field OCTA shows comparable performance to wide-field fluorescein angiography and superior performance to color fundus photography for RNV detection. 48 Even when pathology can be identified in conventional fields of view, wide-field imaging may be necessary to reveal its full extent. An example is ischemia caused by branch retinal vein occlusion 34 or non-perfusion area in diabetic retinopathy (DR). 49 , 50 And finally, we know from other imaging modalities and gene expression profiling that the posterior pole and peripheral retina have different physiological characteristics. 51

High-Resolution OCTA

Transverse and axial resolution in OCTA are decoupled and can be separately optimized. Axial resolution can be important in structural OCT to improve the precision of thickness measurements 52 – 56 but in OCTA we often use en face data representations because the retinal plexuses and choriocapillaris are each laminar tissues oriented largely perpendicular to the scan direction. 57 – 59 Such visualizations are created by projecting over specific tissue layers. For these displays it is important to maintain transverse resolution in both superficial and deep tissue to best capture vascular morphology ( Fig. 3 ).

Иллюстрация — в первоисточнике ↗

Figure 3. High transverse resolution structural OCT and OCTA images of retinal tissue from the vitreous to Haller's layer. ( A ) Cross-sectional image showing the boundaries of the tissue layers and plexuses imaged in the following en face images. ( B ) Macrophage-like cells visualized in the structural channel above the inner limiting membrane. ( C ) Nerve fiber bundles in the nerve fiber layer. ( D – F ) Vessels in the retinal plexuses display distinct morphology. ( G – I ) Choroidal vessels similarly display unique characteristics when imaged at high resolution. With permission from Hormel et al. 34

Transverse resolution can be limited by the anatomy of the eye. Pupil size imposes an upper limit on beam diameter, and, hence, diffraction limited transverse resolution. However, most systems eschew this limit because wider beam diameters are more vulnerable to optical aberrations in the field outside of the central macula. A common choice for OCT has been a 2-mm beam diameter, which gives ∼15-µm transverse resolution. This is actually too low to resolve capillaries at their anatomic width (which can be as small as ∼5 µm 60 ) and is one reason some researchers prefer to skeletonize vessels when calculating vessel density. However, the challenges from wider beam size can be alleviated by real-time OCTA quality display 41 and high sampling density. 31

Aberrations can also be corrected in hardware. The conventional way is with adaptive optics (AO). 61 Early AO-OCTA systems used AO optical setups that included deformable mirrors and wavefront sensors. 62 , 63 The increased cost of these components makes such systems difficult to justify as a translational technology. Sensorless AO-OCTA systems have been demonstrated and provide similar image quality but could provide a better option for clinical adoption 64 or other applications such as murine imaging. 65 It is, however, noteworthy that a similar sensorless AO-OCT system and other AO-OCT devices can observe capillaries without using motion contrast in anterior plexuses; that is, without OCTA processing. 66 This is more difficult in the superficial vascular plexus and in particular, for radial peripapillary capillaries, which have very similar reflectivity to the surrounding static tissue. 67 A purely reflectance signal-based approach also loses the functional information from OCTA- abnormal flow in vessels is not obvious. Another benefit of the high-resolution imaging in these system is that they can avoid projection artifacts, which are a major complication in OCTA imaging that will be discussed below (Section: Artifact Correction).

High Dynamic Range OCTA

Flow signal magnitude in OCTA depends on both intrinsic features of the tissue being imaged—which include erythrocyte density, vessel caliber, and flow velocity—and extrinsic factors, the most important of which is interscan time. Even keeping other intrinsic factors and interscan time fixed, flow signal magnitude is only linear with flow velocity within a limited flow velocity regime outside of which flow is either undetectable (too slow) or saturated (further increase in flow velocity does not increase flow magnitude). 68 In practice what this means is that there isn't a single ideal interscan time because different flow velocity regimes are better imaged by different interscan times.

High-dynamic range OCTA (HDR-OCTA) recognizes this problem and solves it by processing several repeat scans which can be combined to sample at multiple interscan times. 69 Of course, many devices capture more than two scans to construct the OCTA signal. But most simply average the signal across a single interscan time interval Δ t to enhance it. Instead, HDR-OCTA also measures the flow signal between multiples of the interscan time (e.g., 2Δ t , 3Δ t , etc.) to extend the flow detection regime to slower flow. With enough repeats Δ t can also small enough to extend the dynamic range in the opposite direction. An efficient way to implement this approach is by using efficient scanning patterns. OCTA systems usually use raster scanning protocols. But raster scanning is actually inefficient because it wastes time on flyback, during which the system realigns at the edge of a sampling region without collecting data. Bidirectional and interleaved scanning are alternatives that both eliminate flyback time. 70 , 71

Actually, using multiple interscan times can do more than just extend the dynamic range. By comparing flow signal magnitude across multiple interscan times flow speed can also be assessed; this method is called variable interscan time analysis. 72 , 73 Beyond sampling flow at different intervals the multiple interscan times are important for correcting for pulsatility ( Fig. 4 ). 74 Reliable measurement of flow speed can be an important improvement to OCTA imaging since it extends the functional aspect of the technique from an all-or-nothing, flow-or-not visualization to assess pathologically slow/fast flow could be detected.

Иллюстрация — в первоисточнике ↗

Figure 4. Flow speed measurement using variable interscan time analysis (VISTA). ( A ) VISTA images showing the combined superficial vascular complex and intermediate capillary plexus ( top row ) and the deep capillary plexus (DCP) of volunteers with different severities of DR. Flow speeds α are indicated by color ( panel on right ). ( B ) Blow-up of the area within the dotted rectangle in ( A ), showing tortuous capillaries with high α ( triangles ) and focal bulge with low α ( arrow ). With permission from Hwang et al. 74

Handheld OCTA

Pediatric populations provide a different context for OCTA imaging. With infants we cannot use fixation targets or otherwise rely on a cooperative patient, and so we use handheld devices. These devices must balance different design considerations than conventional OCTA devices, including a short working distance and limited procedure times (because babies will begin to cry). The lower limit for working distance is a contact probe, but this should not necessarily be regarded as a limitation for OCTA devices. 75 But shorter procedure times do represent an improvement for the technology. Since the first demonstration of OCTA in a handheld system, 75 the time required to image similar fields of view has dropped; OCTA requiring 3.2 seconds for a 36° has been demonstrated in a handheld device. 76 Swift image acquisition can help reduce motion artifacts, but these can also be addressed separately. 77 Handheld devices also face a unique requirement for ergonomics; for example, OCTA devices specialized for supine imaging have been demonstrated. 78 Finally, similar to conventional OCTA, field of view in handheld devices is increasing. The current state of the art for handheld OCTA is a 140° contact probe system that uses an 800-kHz swept source laser for sampling and a telecentric lens relay system to reduce beam wandering ( Fig. 5 ). 79

Иллюстрация — в первоисточнике ↗

Figure 5. A 140° field of view OCTA images acquired with a handheld device. ( A ) En face OCT image of the right eye from an infant (born at 23 weeks gestation, 593 grams, and imaged at 38 weeks postmenstrual age) with retinopathy of prematurity stage 1, where the fibrovascular ridge is marked by green arrowheads but is not distinctly visible in the structural en face image. ( B ) The corresponding en face OCTA heatmap, which enhances the visualization of the vascularized retina, providing greater clarity and detail. With permission from Ni et al. 75

Advances in OCTA Image Processing and Analysis

In the preceding we discussed improvements in the type and quality of data that OCTA can acquire. Now we want to discuss advances in what we can do with that data.

Artifact Removal

Both visualization and quantification of OCTA data can be improved by addressing artifacts. Unfortunately in OCTA extra care must be taken to correct artifacts relative to structural OCT.

Projection artifacts are a type of artifact unique to OCTA. 79 They duplicate superficial vascular patterns in deeper tissue layers. Visually they disrupt interpretation of vascular morphology in these deeper layers, for example by obscuring the lobular pattern of deep capillary plexus vasculature. On cross-section they create long tails on vessels. Quantitatively they lead to measurements of vascular parameters that should be isolated to specific plexuses to instead be smeared across the target region and superficial flow patterns. Furthermore, they can disrupt detection of macular neovascularization (MNV). 80

Projection artifacts are caused by time-varying shadowing because of flow in overlying vessels and photons being scattered by multiple scatting events. They attenuate with distance from the source vessel. State-of-the-art projection artifact removal can account for this effect to remove artifactual flow signal volumetrically without over-processing removing in situ flow ( Fig. 6 ). 81 This contrasts with early methods for projection artifact removal that both over-processed images of deeper tissue and worked only for en face images rather than volumetrically. 82 – 84

Иллюстрация — в первоисточнике ↗

Figure 6. A normal eye imaged with uncorrected OCTA (row 1), projection-resolved signal attenuation compensated OCTA (sacPR-OCTA, row 2) showing en face OCT angiogram of the superficial vascular plexus (SVC, column A), the intermediate capillary plexus (ICP, column B), the deep capillary plexus (DCP, column C), the outer retina (column D) and the choriocapillaris (CC, column E), and the cross-sectional structural OCT overlaid with color-coded flow signal (column F; violet : inner retina; yellow : outer retina; red : choroid) at the position indicated by the white dotted line in A1. The sacPR-OCTA algorithm removed more artifacts while maintaining vascular integrity, and eliminated artifactual vessel tails (F). With permission from Wang et al. 81

Another way that image quality can be improved is through noise reduction. Classical approaches to this topic typically consist of applying handcrafted filters to images in order to augment specific features. For example, vesselness filters can be applied to en face OCTA angiograms to emphasize vascular patterns. 85 Alternatively, background can also be reduced. In OCTA bulk motion is a large contributor to background, so in this case an example is an iterative approach to reducing its signal strength. 86 – 88 Statistical approaches that attempt to estimate flow signal value are also effective and can be applied to deeper layers where signal attenuation is a factor. 89

A disadvantage to handcrafted algorithms like these is that they lack context awareness that can be trained into neural networks. This is a clear advantage for deep learning-based approaches. For denoising the subtlety is in choosing denoised images: how do we select cleaner images for networks to learn from? One option is train networks to reconstruct high-definition scans from low definition, undersampled scans; interestingly, networks trained in this way were actually able to improve scan quality in high-definition scans, as well as undersampled ones ( Fig. 7 ). 90 , 91 Another approach is to train networks to improve angiograms constructed with a lower number of repeats from angiograms with more repeats, because the higher number of repeats improves signal quality. 35 , 92

Иллюстрация — в первоисточнике ↗

Figure 7. Artifact-free OCTA image enhancement using deep learning. Top row: raw images; bottom row: images denoised with a deep learning network. Note that this method did not produce artifactual vessels, as can occur with handcrafted vesselness filters. With permission from Gao et al. 90

Deep Learning in Feature Detection and Segmentation

Quantification of OCTA angiograms can stage and diagnose prevalent vision threatening diseases. 93 , 94 Originally, OCTA quantification focused on vessel density 95 – 99 or vessel morphology metrics such as tortuosity or fractal dimension. 100 These approaches have been reviewed elsewhere. 80 , 93 An alternative proxy to vessel density for perfusion loss is non-perfusion area (NPA), which is defined as vessel-free regions exceeding an area threshold. 101 , 102 An advantage for NPA over vessel density is that it can provide a visual cue for the location of perfusion loss, which can be useful for assessing disease processes ( Fig. 8 ). With the aid of deep learning NPA can be quantified in separately in individual plexuses 103 and wide-field images. 104 Such measurements correlate with retinopathy 105 and glaucoma severity. 106 – 110

Иллюстрация — в первоисточнике ↗

Figure 8. Automated NPA segmentation results on montaged wide-field OCT angiography imaging for an eye with severe nonproliferative DR. ( A ) Angiogram image; ( B ) angiogram with labeled NPA ( teal ) and shadow artifacts ( yellow ). By avoiding detecting shadow as NPA, this deep learning model improves NPA quantification. With permission from Guo et al. 104

Vasculature in the posterior segment is perhaps the most intricate in the human body. As such, not all measurements are suitable in every anatomic region. Although NPA is useful for assessing perfusion in the retinal plexuses, it is not well suited for assessing perfusion defects in the choriocapillaris where vasculature is dense, and perfusion can be lost without correspondingly large vessel-free regions. Instead, perfusion loss in the choriocapillaris is best assessed by detecting flow deficits. This task is difficult because vessel density in the choriocapillaris is high enough that individual capillaries are difficult to resolve, and, because of its location posterior to the retina, requires imaging in a region necessarily affected by signal attenuation and possibly complicated by overlying pathology such as drusen or retinal fluid. 111 But if it is difficult to measure flow deficits, it is also useful—flow deficit quantification can predict which late-stage form of age-related macular degeneration is likely to develop 112 and predict how rapidly the hallmark features for these stages (macular neovascularization and geographic atrophy) will develop 113 – 116 and where. 117 Flow deficits also correlate with disease severity in DR. 118

MNV is itself an excellent target for OCTA, but visualization is complicated by the posterior location of the pathology. It is best analyzed with projection artifact removal. 81 , 119 With projection artifact removal OCTA is a powerful tool for interrogating MNV. As a volumetric imaging modality OCTA can distinguish MNV type, 94 and artificial intelligence can be used automatically to detect and segment lesions ( Fig. 9 ), 120 , 121 enabling efficient monitoring of lesion development and treatment response. 122 – 124

Иллюстрация — в первоисточнике ↗

Figure 9. MNV membrane segmentation on 3 × 3-mm scans. ( A ) Inner retinal angiogram. ( B ) Projection-resolved outer retinal angiogram with MNV membrane ground truth ( white outline ). ( C ) MNV probability map. (D) Structural OCT cross section at the location of the white line in ( A ), with flow signal overlaid ( violet : inner retinal; yellow : pathologic outer retinal; red : choroidal). Case 1 shows a large MNV lesion that cannot be fully imaged using a small scanning window. Case 2 shows a small MNV lesion in a defocused scan with strong residual artifacts. These cases illustrate the versatility of an deep learning based approach for MNV segmentation. With permission from Wang et al. 121

Both MNV and NPA represent features with mature image analytic approaches already developed for segmentation and detection. There are other features OCTA can address. Microaneurysms can be detected and categorized by appearance, transverse and axial location, and the presence/amount of flow signal using OCTA. 125 – 130 The last two capabilities stand in contrast to dye-based angiography, which remains useful for detecting microaneurysms because it can image the periphery. 131 , 132 (Although as discussed above contemporary OCTA devices are incorporating larger fields of view.) 34 – 36 Another pathology that OCTA is excellent for identifying that benefits from larger fields of view is retinal neovascularization. OCTA is superior for retinal neovascularization detection in comparison to dye-based angiography because it can readily identify neovascular vessels due to their location above the internal limiting membrane. 46 , 133 But similar to microaneurysms, retinal neovascularization detection should really be performed on wide-field systems to avoid missing peripheral pathology. 45 , 48 Other examples of pathology in which OCTA provides superior detection capabilities compared to allied retinal imaging modalities (fundus photography, dye-based angiography) include collateral vessels 134 – 136 and intraretinal microvascular abnormalities. 137 – 139

Many of the features just described await automated or precise quantification methods based on artificial intelligence. But there are other features artificial intelligence can be used to infer and characterize that are difficult for human graders to recognize. One example is differentiating arteries and veins. Although time consuming by visual inspection, deep learning networks can accurately differentiate between these vessel types. 140 – 143 Another “difficult-for-humans” task artificial intelligence can accomplish is to infer which vessels would appear leaky under dye-based angiography. 144 This capability is useful because it could enable detection of pathological vessels without the need for an invasive procedure, and because vessel leakage obscures vasculature in dye angiography. Both these drawbacks could be avoided with an OCTA-based approach.

Deep Learning in Diagnostics

Feature quantification is useful for disease diagnostics, prognostics, and investigations of etiology and pathophysiology. But in clinical practice we are often interested in screening, which can be done directly from an image without feature quantification using deep learning. For example, networks trained to diagnose DR directly from OCTA images have demonstrated human level performance. 145 – 150 Similarly, DR-related pathology (diabetic macular ischemia) is also amenable to AI-based screening. 151 Probably most effort has addressed DR because it is an obvious target for OCTA given the microvascular nature of the disease, but diagnostic networks exist for other prominent diseases including glaucoma 152 and age-related macular degeneration. 153 Because having one disease doesn't mean you can't have another, diagnosing multiple diseases simultaneously can be an efficient use of resources. 153

Unfortunately, there is a relatively small number of publications incorporating OCTA for diagnostics; many networks get by with just structural OCT data. 154 – 156 Given that many retinal diseases include a vascular component, this may be an oversight. For example, even in the case of a conventionally structural pathology (DME), OCTA data can offer quantitative advantage. 157 Collection of larger OCTA datasets is probably the easiest way to address this limitation.

An important issue for any diagnostic network is explainability. To the best of our knowledge this has not been explored in an OCTA context very thoroughly. However, explainability approaches that do incorporate OCTA information can help interpret of AI-based output ( Fig. 10 ). 158

Иллюстрация — в первоисточнике ↗

Figure 10. An explainability approach, biomarker activation maps (BAMs), selectively highlights clinically relevant information. Here, an example is shown for a correctly predicted referable DR scan. ( A ) Raw OCTA angiogram of the superficial vascular plexus. ( B ) Explainability framework output, showing pixels the network considered pathologic (with pathologic severity increasing from blue to yellow ). ( C ) An NPA segmentation network output showing NPA in teal . Note the correspondence between the explainability frameworks highlighted pixels and the distribution of NPA.

Future Directions

Perhaps the most important trends in OCTA technology can be regarded as improved acquisition speed and improved image analysis. For the latter, artificial intelligence–based approaches are playing a leading role. An important example is retinal layer segmentation, which is often the bottleneck in clinical OCTA throughput. Accurate layer segmentation is also critical for OCTA quantification, where context sensitivity for pathologies like retinal fluid is a major advantage for AI-based approaches. But these approaches rely on large datasets from which to train models. Large public datasets are not yet available for the largest fields of view, high-resolution images (in particular of deeper tissue layers), or for more specialized applications like images measuring flow speed. As these datasets are developed more potent image analysis can also be achieved. For the former (instrumentation), improved acquisition speeds opened the way to new types of measurements (flow speed) and detection sensitivity and dynamic range. And finer resolution and larger fields of view mean we can measure vascular anatomy more extensively. Because of these trends we speculate that OCTA will be increasingly used as a clinical and research technology.

Summary

Conventional limitations to OCTA technology such as field of view, flow detection dynamic range, and imaging artifacts can now be elided. We can expect instruments to continue to improve as hardware does, and image analysis techniques to improve as more powerful machine learning algorithms are introduced.

Acknowledgments

Supported by grants from National Institutes of Health (R01 EY 036429, R01 EY035410, R01 EY024544, R01 EY027833, R01 EY031394, R43EY036781, P30 EY010572, T32 EY023211, UL1TR002369); the Jennie P. Weeks Endowed Fund; the Malcolm M. Marquis, MD Endowed Fund for Innovation; Unrestricted Departmental Funding Grant and H. James and Carole Free Catalyst Award from Research to Prevent Blindness (New York, NY); Edward N. & Della L. Thome Memorial Foundation Award, and the Bright Focus Foundation (G2020168, M20230081).

Disclosure: T.T. Hormel , IFOCUS Imaging (I); D. Huang , Boehringer Ingelheim (C), Cylite (F), Elsevier (R), Genentech (P, R), Intalight (F), Kugler (R), Visionix (F, P, R, C); Y. Jia , IFOCUS Imaging (I), Optovue/Visionix, Inc. (P, R), Genentech (P, R, F), Optos Inc. (P)

References

1. Huang D , Swanson EA , Lin CP , et al . Optical coherence tomography . Science (1979) . 1991 ; 22 ( 2545035 ): 1178 – 1181 . 10.1126/science.1957169 PMC4638169 1957169

2. Chen TC , Cense B , Miller JW , et al . Histologic correlation of in vivo optical coherence tomography images of the human retina . Am J Ophthalmol . 2006 ; 141 : 1165 – 1168 . 16765704 10.1016/j.ajo.2006.01.086

3. Flaxel CJ , Adelman RA , Bailey ST , et al . Age-related macular degeneration preferred practice pattern . Ophthalmology . 2020 ; 127 ( 1 ): P1 – P65 . 31757502 10.1016/j.ophtha.2019.09.024

4. Flaxel CJ , Adelman RA , Bailey ST , et al . Diabetic retinopathy preferred practice pattern . Ophthalmology . 2020 ; 127 ( 1 ): P66 – P145 . 31757498 10.1016/j.ophtha.2019.09.025

5. Gonzalez VH , Campbell J , Holekamp NM , et al . Early and long-term responses to anti–vascular endothelial growth factor therapy in diabetic macular edema: analysis of Protocol I Data . Am J Ophthalmol . 2016 ; 172 : 72 – 79 . 27644589 10.1016/j.ajo.2016.09.012

6. Lee CS , Tyring AJ , Wu Y , et al . Generating retinal flow maps from structural optical coherence tomography with artificial intelligence . Sci Rep . 2019 ; 9 : 1 – 11 . 30952891 10.1038/s41598-019-42042-y PMC6450899

7. Olver JM. Functional anatomy of the choroidal circulation: methyl methacrylate casting of human choroid . Eye (Basingstoke) . 1990 ; 4 : 262 – 272 . 10.1038/eye.1990.38 2379644

8. Iadecola C. The neurovascular unit coming of age: a journey through neurovascular coupling in health and disease . Neuron . 2017 ; 96 : 17 – 42 . 28957666 10.1016/j.neuron.2017.07.030 PMC5657612

9. Patel RC , Wang J , Hwang TS , et al . Plexus-specific detection of retinal vascular pathologic conditions with projection-resolved OCT angiography . Ophthalmol Retina . 2018 ; 2 : 816 – 826 . 30148244 10.1016/j.oret.2017.11.010 PMC6106769

10. Hee MR , Izatt JA , Swanson EA , et al . Optical coherence tomography of the human retina . Arch Ophthalmol . 1995 ; 113 : 325 – 332 . 7887846 10.1001/archopht.1995.01100030081025

11. Wojtkowski M , Srinivasan VJ , Ko TH , Fujimoto JG , Kowalczyk A , Duker JS. Ultrahigh-resolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensation . Optics Exp . 2004 ; 12 : 2404 – 2422 . 10.1364/opex.12.002404 19475077

12. Fercher AF , Hitzenberger CK , Kamp G , El-Zaiat SY . Measurement of intraocular distances by backscattering spectral interferometry . Opt Commun . 1995 ; 117 ( 1-2 ): 43 – 48 .

13. Nassif N , Cense B , Hyle Park B , et al . In vivo human retinal imaging by ultrahigh-speed spectral domain optical coherence tomography . Opt Lett . 2004 ; 29 : 480 . 15005199 10.1364/ol.29.000480

14. Choma MA , Sarunic M V , Yang C , Izatt JA. Sensitivity advantage of swept source and Fourier domain optical coherence tomography . Opt Express . 2003 ; 11 : 2183 – 2189 . 19466106 10.1364/oe.11.002183

15. Fingler J , Schwartz D , Yang C , Fraser SE. Mobility and transverse flow visualization using phase variance contrast with spectral domain optical coherence tomography . Opt Express . 2007 ; 15 : 12636 – 12653 . 19550532 10.1364/oe.15.012636

16. Fingler J , Zawadzki RJ , Werner JS , Schwartz D , Fraser SE. Volumetric microvascular imaging of human retina using optical coherence tomography with a novel motion contrast technique . Opt Express . 2009 ; 17 : 22190 – 22200 . 19997465 10.1364/OE.17.022190 PMC2791341

17. Makita S , Hong Y , Yamanari M , Yatagai T , Yasuno Y. Optical coherence angiography . Opt Express . 2006 ; 14 : 7821 – 7840 . 19529151 10.1364/oe.14.007821

18. An L , Wang RK . In vivo volumetric imaging of vascular perfusion within human retina and choroids with optical micro-angiography . Opt Express . 2008 ; 16 : 11438 – 11452 . 18648464 10.1364/oe.16.011438

19. Jia Y , Tan O , Tokayer J , et al . Split-spectrum amplitude-decorrelation angiography with optical coherence tomography . Opt Express . 2012 ; 20 : 4710 – 4725 . 22418228 10.1364/OE.20.004710 PMC3381646

20. Chinn SR , Swanson EA , Fujimoto JG. Optical coherence tomography using a frequency-tunable optical source . Opt Lett . 1997 ; 22 : 340 . 18183195 10.1364/ol.22.000340

21. Huber R , Wojtkowski M , Fujimoto JG. Fourier domain mode locking (FDML): a new laser operating regime and applications for optical coherence tomography . Opt Express . 2006 ; 254 : 3225 – 3237 . 10.1364/oe.14.003225 19516464

22. Häusler G , Lindner MW. “Coherence radar” and “spectral radar”— new tools for dermatological diagnosis . J Biomed Opt . 1998 ; 3 : 21 – 31 . 23015002 10.1117/1.429899

23. Wojtkowski M , Leitgeb R , Kowalczyk A , Bajraszewski T , Fercher AF. In vivo human retinal imaging by Fourier domain optical coherence tomography . J Biomed Opt . 2002 ; 7 : 457 . 12175297 10.1117/1.1482379

24. Gaur P , Grieco A , Fainman Y. Multirate spectral domain optical coherence tomography . IEEE Photonics J . 2023 ; 15 ( 5 ): 1 – 10 .

25. Zhang J , Nguyen T , Potsaid B , et al . Multi-MHz MEMS-VCSEL swept-source optical coherence tomography for endoscopic structural and angiographic imaging with miniaturized brushless motor probes . Biomed Opt Express . 2021 ; 12 : 2384 . 33996236 10.1364/BOE.420394 PMC8086463

26. Kim TS , Joo JY , Shin I , et al . 9.4 MHz A-line rate optical coherence tomography at 1300 nm using a wavelength-swept laser based on stretched-pulse active mode-locking . Sci Rep . 2020 ; 10 : 9328 . 32518256 10.1038/s41598-020-66322-0 PMC7283258

27. Chen S , Potsaid B , Li Y , et al . High speed, long range, deep penetration swept source OCT for structural and angiographic imaging of the anterior eye . Sci Rep . 2022 ; 12 : 992 . 35046423 10.1038/s41598-022-04784-0 PMC8770693

28. Zhang Q , Lee CS , Chao J , et al . Wide-field optical coherence tomography based microangiography for retinal imaging . Sci Rep . 2016 ; 6 ( 1 ): 22017 . 26912261 10.1038/srep22017 PMC4766473

29. Couturier A , Rey PA , Erginay A , et al . Widefield OCT-angiography and fluorescein angiography assessments of nonperfusion in diabetic retinopathy and edema treated with anti–vascular endothelial growth factor . Ophthalmology . 2019 ; 126 : 1685 – 1694 . 31383483 10.1016/j.ophtha.2019.06.022

30. Mase T , Ishibazawa A , Nagaoka T , Yokota H , Yoshida A. Radial peripapillary capillary network visualized using wide-field montage optical coherence tomography angiography . Invest Ophthalmol Vis Sci . 2016 ; 57 : 504 – 510 . 10.1167/iovs.15-18877 27454659

31. Wei X , Hormel TT , Guo Y , Jia Y. 75-degree non-mydriatic single-volume optical coherence tomographic angiography . Biomed Opt Express . 2019 ; 10 : 6286 – 6295 . 31853400 10.1364/BOE.10.006286 PMC6913399

32. Wei X , Hormel TT , Guo Y , Hwang TS , Jia Y. High-resolution wide-field OCT angiography with a self-navigation method to correct microsaccades and blinks . Biomed Opt Express . 2020 ; 11 : 3234 – 3245 . 32637251 10.1364/BOE.390430 PMC7316026

33. Wei X , Hormel TT , Jia Y. Phase-stabilized complex-decorrelation angiography . Biomed Opt Express . 2021 ; 12 : 2419 – 2431 . 33996238 10.1364/BOE.420503 PMC8086438

34. Hormel TT , Liang GB , Wei X , et al . Visualizing features with wide-field volumetric OCT angiography . Opt Express . 2024 ; 32 : 10329 . 38571248 10.1364/OE.510640 PMC11018334

35. Niederleithner M , De Sisternes L , Stino H , et al . Ultra-widefield OCT angiography . IEEE Trans Med Imaging . 2023 ; 42 : 1009 – 1020 . 36383595 10.1109/TMI.2022.3222638

36. Liang GB , Hormel TT , Wei X , et al . Single-shot OCT and OCT angiography for slab-specific detection of diabetic retinopathy . Biomed Opt Express . 2023 ; 14 : 5682 . 38021127 10.1364/BOE.503476 PMC10659794

37. Mehta N , Cheng Y , Yasin Alibhai A , Duker JS , Wang RK , Waheed NK . Optical coherence tomography angiography distortion correction in widefield montage images . Quant Imaging Med Surg . 2021 ; 11 : 928 – 938 . 33654666 10.21037/qims-20-791 PMC7829171

38. Vienola K V. , Braaf B , Sheehy CK , et al . Real-time eye motion compensation for OCT imaging with tracking SLO . Biomed Opt Express . 2012 ; 3 : 2950 . 23162731 10.1364/BOE.3.002950 PMC3493227

39. Zhang Q , Huang Y , Zhang T , et al . Wide-field imaging of retinal vasculature using optical coherence tomography-based microangiography provided by motion tracking . J Biomed Opt . 2015 ; 20 ( 6 ): 066008 . 26102573 10.1117/1.JBO.20.6.066008 PMC4478052

40. Braaf B , Vienola K V. , Sheehy CK , et al . Real-time eye motion correction in phase-resolved OCT angiography with tracking SLO . Biomed Opt Express . 2013 ; 4 : 51 . 23304647 10.1364/BOE.4.000051 PMC3539196

41. Wei X , Camino A , Pi S , et al . Real-time cross-sectional and en face OCT angiography guiding high-quality scan acquisition . Opt Lett . 2019 ; 44 : 1431 . 30874667 10.1364/OL.44.001431 PMC7188388

42. Kim HJ , Song BJ , Choi Y , Kim BM. Cross-scanning optical coherence tomography angiography for eye motion correction . J Biophotonics . 2020 ; 13 ( 9 ): e202000170 . 32475032 10.1002/jbio.202000170

43. Ploner SB , Kraus MF , Moult EM , et al . Efficient and high accuracy 3-D OCT angiography motion correction in pathology . Biomed Opt Express . 2021 ; 12 : 125 . 33520381 10.1364/BOE.411117 PMC7818965

44. Ni S , Wei X , Ng R , et al . High-speed and widefield handheld swept-source OCT angiography with a VCSEL light source . Biomed Opt Express . 2021 ; 12 : 3553 . 34221678 10.1364/BOE.425411 PMC8221946

45. You QS , Guo Y , Wang J , et al . Detection of clinically unsuspected retinal neovascularization with wide-field optical coherence tomography angiography . Retina . 2019 ; 00 ( 00 ): 1 . 10.1097/IAE.0000000000002487 PMC6722037 30845022

46. Tsuboi K , Mazloumi M , Guo Y , et al . Utility of en face OCT for the detection of clinically unsuspected retinal neovascularization in patients with diabetic retinopathy . Ophthalmol Retina . 2023 ;( Ilm ): 1 – 9 . 36918122 10.1016/j.oret.2023.03.002 PMC10440281

47. Sawada O , Ichiyama Y , Obata S , et al . Comparison between wide-angle OCT angiography and ultra-wide field fluorescein angiography for detecting non-perfusion areas and retinal neovascularization in eyes with diabetic retinopathy . Graefes Arch Clin Exp Ophthalmol . 2018 ; 256 : 1275 – 1280 . 29713816 10.1007/s00417-018-3992-y

48. Cui Y , Zhu Y , Wang JC , et al . Comparison of widefield swept-source optical coherence tomography angiography with ultra-widefield colour fundus photography and fluorescein angiography for detection of lesions in diabetic retinopathy . Br J Ophthalmol . 2021 ; 105 : 577 – 581 . 32591347 10.1136/bjophthalmol-2020-316245 PMC9092310

49. Alibhai AY , De Pretto LR , Moult EM , et al . Quantification of retinal capillary nonperfusion in diabetics using wide-field optical coherence tomography angiography . Retina . 2020 ; 40 : 412 – 420 . 30576300 10.1097/IAE.0000000000002403

50. Stino H , Huber KL , Niederleithner M , et al . Association of diabetic lesions and retinal nonperfusion using widefield multimodal imaging . Ophthalmol Retina . 2023 ; 7 : 1042 – 1050 . 37517798 10.1016/j.oret.2023.07.020

51. Quinn N , Csincsik L , Flynn E , et al . The clinical relevance of visualising the peripheral retina . Prog Retin Eye Res . 2019 ; 68 : 83 – 109 . 30316018 10.1016/j.preteyeres.2018.10.001

52. Diabetic retinopathy clinical research network. relationship between optical coherence tomography-measured central retinal thickness and visual acuity in diabetic macular edema . Ophthalmology . 2007 ; 114 : 525 – 536 . 17123615 10.1016/j.ophtha.2006.06.052 PMC2585542

53. Ajtony C , Balla Z , Somoskeoy S , Kovacs B. Relationship between visual field sensitivity and retinal nerve fiber layer thickness as measured by optical coherence tomography . Invest Ophthalmol Vis Sci . 2007 ; 48 : 258 – 263 . 17197541 10.1167/iovs.06-0410

54. Jia Y , Simonett JM , Wang J , et al . Wide-field OCT angiography investigation of the relationship between radial peripapillary capillary plexus density and nerve fiber layer thickness . Invest Ophthalmol Vis Sci . 2017 ; 58 : 5188 – 5194 . 29049718 10.1167/iovs.17-22593 PMC5637456

55. You QS , Tsuboi K , Guo Y , et al . Comparison of central macular fluid volume with central subfield thickness in patients with diabetic macular edema using optical coherence tomography angiography . JAMA Ophthalmol . 2021 ; 139 : 734 – 741 . 33983385 10.1001/jamaophthalmol.2021.1275 PMC8120439

56. Hood DC , Fortune B , Mavrommatis MA , et al . Details of glaucomatous damage are better seen on OCT en face images than on OCT retinal nerve fiber layer thickness maps . Invest Ophthalmol Vis Sci . 2015 ; 56 : 6208 – 6216 . 26426403 10.1167/iovs.15-17259 PMC4703406

57. Snodderly D , Weinhaus R , Choi J. Neural-vascular relationships in central retina of macaque monkeys (Macaca fascicularis) . J Neuroscience . 2018 ; 12 : 1169 – 1193 . 10.1523/JNEUROSCI.12-04-01169.1992 PMC6575794 1556592

58. Campbell JP , Zhang M , Hwang TS , et al . Detailed vascular anatomy of the human retina by projection-resolved optical coherence tomography angiography . Sci Rep . 2017 ; 7 ( February ): 1 – 11 . 28186181 10.1038/srep42201 PMC5301488

59. Nesper PL , Fawzi AA. Human parafoveal capillary vascular anatomy and connectivity revealed by optical coherence tomography angiography . Invest Ophthalmol Vis Sci . 2018 ; 59 : 3858 – 3867 . 30073360 10.1167/iovs.18-24710 PMC6071478

60. Shimizu K , Ujiie K. Structure of Ocular Vessels . Tokyo : Igaku-Shoin ; 1978 .

61. Babcock HW. The possibility of compensating astronomical seeing . Publications Astronom Soc Pac . 1953 ; 65 ( 386 ): 229 – 236 .

62. Kurokawa K , Sasaki K , Makita S , Hong YJ , Yasuno Y. Three-dimensional retinal and choroidal capillary imaging by power Doppler optical coherence angiography with adaptive optics . Opt Express . 2012 ; 20 : 22796 – 22812 . 23037430 10.1364/OE.20.022796

63. Salas M , Augustin M , Ginner L , et al . Visualization of micro-capillaries using optical coherence tomography angiography with and without adaptive optics . Biomed Opt Express . 2017 ; 8 : 207 . 28101412 10.1364/BOE.8.000207 PMC5231293

64. Camino A , Zang P , Athwal A , et al . Sensorless adaptive-optics optical coherence tomographic angiography . Biomed Opt Express . 2020 ; 11 : 3952 – 3967 . 33014578 10.1364/BOE.396829 PMC7510908

65. Wei X , Hormel TT , Pi S , Wang B , Morrison JC , Jia Y. Wide-field sensorless adaptive optics swept-source optical coherence tomographic angiography in rodents . Opt Lett . 2022 ; 47 : 5060 . 36181186 10.1364/OL.472387

66. Ju MJ , Heisler M , Wahl D , Jian Y , Sarunic M V . Multiscale sensorless adaptive optics OCT angiography system for in vivo human retinal imaging . J Biomed Opt . 2017 ; 22 ( 12 ): 1 . 10.1117/1.JBO.22.12.121703 29094524

67. Kurokawa K , Nemeth M. Multifunctional adaptive optics optical coherence tomography allows cellular scale reflectometry, polarimetry, and angiography in the living human eye . Biomed Opt Express . 2024 ; 15 : 1331 . 38404344 10.1364/BOE.505395 PMC10890865

68. Su JP , Chandwani R , Gao SS , et al . Calibration of optical coherence tomography angiography with a microfluidic chip . J Biomed Opt . 2016 ; 21 ( 8 ): 086015 . 27557344 10.1117/1.JBO.21.8.086015 PMC4995373

69. Wei X , Hormel TT , Pi S , Guo Y , Jian Y , Jia Y. High dynamic range optical coherence tomography angiography (HDR-OCTA) . Biomed Opt Express . 2019 ; 10 : 3560 . 31360605 10.1364/BOE.10.003560 PMC6640830

70. Wei X , Camino A , Pi S , et al . A novel and effective scan pattern for velocimetric OCT angiography . In: Izatt JA , Fujimoto JG , eds. Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XXIII . Bellingham, WA: SPIE ; 2019 ;( 10867 ): 20 .

71. Ju MJ , Heisler M , Athwal A , Sarunic M V. , Jian Y. Effective bidirectional scanning pattern for optical coherence tomography angiography . Biomed Opt Express . 2018 ; 9 : 2336 . 29760992 10.1364/BOE.9.002336 PMC5946793

72. Arya M , Filho MB , Rebhun CB , et al . Analyzing relative flow speeds in diabetic retinopathy using variable interscan time analysis OCT angiography . Ophthalmol Retina . 2021 ; 5 : 49 – 59 . 32585373 10.1016/j.oret.2020.06.024 PMC8906440

73. Rebhun CB , Moult EM , Novais EA , et al . Polypoidal choroidal vasculopathy on swept-source optical coherence tomography angiography with variable interscan time analysis . Transl Vis Sci Technol . 2017 ; 6 ( 6 ): 4 . 10.1167/tvst.6.6.4 PMC5678546 29134135

74. Hwang Y , Won J , Yaghy A , et al . Retinal blood flow speed quantification at the capillary level using temporal autocorrelation fitting OCTA [Invited] . Biomed Opt Express . 2023 ; 14 : 2658 . 37342704 10.1364/BOE.488103 PMC10278638

75. Yang J , Liu L , Campbell JP , Huang D , Liu G. Handheld optical coherence tomography angiography . Biomed Opt Express . 2017 ; 8 : 2287 . 28736672 10.1364/BOE.8.002287 PMC5516829

76. Song S , Zhou K , Xu JJ , Zhang Q , Lyu S , Wang R. Development of a clinical prototype of a miniature hand-held optical coherence tomography probe for prematurity and pediatric ophthalmic imaging . Biomed Opt Express . 2019 ; 10 : 2383 . 31143494 10.1364/BOE.10.002383 PMC6524597

77. Malone JD , El-Haddad MT , Yerramreddy SS , Oguz I , Tao YKK. Handheld spectrally encoded coherence tomography and reflectometry for motion-corrected ophthalmic optical coherence tomography and optical coherence tomography angiography . Neurophotonics . 2019 ; 6 ( 04 ): 1 . 10.1117/1.NPh.6.4.041102 PMC6991137 32042852

78. Viehland C , Chen X , Tran-Viet D , et al . Ergonomic handheld OCT angiography probe optimized for pediatric and supine imaging . Biomed Opt Express . 2019 ; 10 : 2623 . 31143506 10.1364/BOE.10.002623 PMC6524583

79. Ni S , Ben Liang G , Ng R , et al . Panretinal handheld OCT angiography for pediatric retinal imaging . Biomed Opt Express . 2024 ; 15 : 3412 . 38855676 10.1364/BOE.520739 PMC11161374

80. Hormel TT , Jia Y , Jian Y , et al . Plexus-specific retinal vascular anatomy and pathologies as seen by projection-resolved optical coherence tomographic angiography . Prog Retin Eye Res . 2021 ; 80 ( July 2020 ): 100878 . 32712135 10.1016/j.preteyeres.2020.100878 PMC7855241

81. Wang J , Hormel TT , Bailey ST , Hwang TS , Huang D , Jia Y. Signal attenuation-compensated projection-resolved OCT angiography . Biomed Opt Express . 2023 ; 14 : 2040 – 2054 . 37206138 10.1364/BOE.483835 PMC10191650

82. Jia Y , Bailey ST , Hwang TS , et al . Quantitative optical coherence tomography angiography of vascular abnormalities in the living human eye . Proc Natl Acad Sci USA . 2015 ; 112 ( 18 ): E2395 – E2402 . 25897021 10.1073/pnas.1500185112 PMC4426471

83. Jia Y , Bailey ST , Wilson DJ , et al . Quantitative optical coherence tomography angiography of choroidal neovascularization in age-related macular degeneration . Ophthalmology . 2014 ; 121 : 1435 – 1444 . 24679442 10.1016/j.ophtha.2014.01.034 PMC4082740

84. Zhang A , Zhang Q , Wang RK. Minimizing projection artifacts for accurate presentation of choroidal neovascularization in OCT micro-angiography . Biomed Opt Express . 2015 ; 6 : 4130 – 4143 . 26504660 10.1364/BOE.6.004130 PMC4605069

85. Frangi AF , Niessen WJ , Vincken KL , Viergever MA. Multiscale vessel enhancement filtering . In: Medical Image Computing and Computer-Assisted Intervention . Berlin : Springer ; 1998 ; 1496 : 130 – 137 .

86. Camino A , Jia Y , Liu G , Wang J , Huang D. Regression-based algorithm for bulk motion subtraction in optical coherence tomography angiography . Biomed Opt Express . 2017 ; 8 : 3053 – 3066 . 28663926 10.1364/BOE.8.003053 PMC5480449

87. Camino A , Zhang M , Liu L , Wang J , Jia Y , Huang D. Enhanced quantification of retinal perfusion by improved discrimination of blood flow from bulk motion signal in OCTA . Transl Vis Sci Technol . 2018 ; 7 ( 6 ): 20 , doi: 10.1167/tvst.7.6.20 . PMC6284469 30564509

88. Wei X , Camino A , Pi S , et al . Fast and robust standard-deviation-based method for bulk motion compensation in phase-based functional OCT . Opt Lett . 2018 ; 43 : 2204 – 2207 . 29714790 10.1364/OL.43.002204 PMC6411296

89. Husvogt L , Ploner SB , Chen S , et al . Maximum a posteriori signal recovery for optical coherence tomography angiography image generation and denoising . Biomed Opt Express . 2021 ; 12 : 55 . 33520377 10.1364/BOE.408903 PMC7818957

90. Gao M , Hormel TT , Wang J , et al . An open-source deep learning network for reconstruction of high-resolution oct angiograms of retinal intermediate and deep capillary plexuses . Transl Vis Sci Technol . 2021 ; 10 ( 13 ): 1 – 14 . 10.1167/tvst.10.13.13 PMC8590160 34757393

91. Gao M , Guo Y , Hormel TT , Sun J , Hwang TS , Jia Y. Reconstruction of high-resolution 6×6-mm OCT angiograms using deep learning . Biomed Opt Express . 2020 ; 11 : 3585 . 33014553 10.1364/BOE.394301 PMC7510902

92. Abu-Qamar O , Lewis W , Mendonca LSM , et al . Pseudoaveraging for denoising of OCT angiography: a deep learning approach for image quality enhancement in healthy and diabetic eyes . Int J Retina Vitreous . 2023 ; 9 ( 1 ): 62 . 37822004 10.1186/s40942-023-00486-5 PMC10568842

93. Hormel TT , Jia Y. OCT angiography and its retinal biomarkers [Invited] . Biomed Opt Express . 2023 ; 14 : 4542 – 4566 . 37791289 10.1364/BOE.495627 PMC10545210

94. Hormel TT , Hwang TS , Bailey ST , Wilson DJ , Huang D , Jia Y. Artificial intelligence in OCT angiography . Prog Retin Eye Res . 2021 ; 85 : 100965 . 33766775 10.1016/j.preteyeres.2021.100965 PMC8455727

95. Kim AY , Chu Z , Shahidzadeh A , Wang RK , Puliafito CA , Kashani AH. Quantifying microvascular density and morphology in diabetic retinopathy using spectral-domain optical coherence tomography angiography . Invest Ophthalmol Vis Sci . 2016 ; 57 : 362 – 370 . 10.1167/iovs.15-18904 PMC4968771 27409494

96. Agemy SA , Scripsema NK , Shah CM , et al . Retinal vascular perfusion density mapping using optical coherence tomography angiography in normals and diabetic retinopathy patients . Retina . 2015 ; 35 : 2353 – 2363 . 26465617 10.1097/IAE.0000000000000862

97. Nishida T , Moghimi S , Wu JH , et al . Association of Initial Optical Coherence Tomography Angiography Vessel Density Loss With Faster Visual Field Loss in Glaucoma . JAMA Ophthalmol . 2022 ; 140 : 319 – 326 . 35201270 10.1001/jamaophthalmol.2021.6433 PMC8874898

98. Bowd C , Belghith A , Zangwill LM , et al . Deep learning image analysis of optical coherence tomography angiography measured vessel density improves classification of healthy and glaucoma eyes . Am J Ophthalmol . 2022 ; 236 : 298 – 308 . 34780803 10.1016/j.ajo.2021.11.008 PMC10042115

99. Rabiolo A , Gelormini F , Sacconi R , et al . Comparison of methods to quantify macular and peripapillary vessel density in optical coherence tomography angiography . PLoS One . 2018 ; 13 ( 10 ): e0205773 . 30335815 10.1371/journal.pone.0205773 PMC6193681

100. Corvi F , Pellegrini M , Erba S , Cozzi M , Staurenghi G , Giani A. Reproducibility of vessel density, fractal dimension, and foveal avascular zone using 7 different optical coherence tomography angiography devices . Am J Ophthalmol . 2018 ; 186 : 25 – 31 . 29169882 10.1016/j.ajo.2017.11.011

101. Hwang TS , Gao SS , Liu L , et al . Automated quantification of capillary nonperfusion using optical coherence tomography angiography in diabetic retinopathy . JAMA Ophthalmol . 2016 ; 134 : 367 – 373 . 26795548 10.1001/jamaophthalmol.2015.5658 PMC4978127

102. Chen S , Moult EM , Zangwill LM , Weinreb RN , Fujimoto JG. Geometric perfusion deficits: a novel OCT angiography biomarker for diabetic retinopathy based on oxygen diffusion . Am J Ophthalmol . 2021 ; 222 : 256 – 270 . 32918905 10.1016/j.ajo.2020.09.007 PMC8015788

103. Wang J , Hormel TT , You Q , et al . Robust non-perfusion area detection in three retinal plexuses using convolutional neural network in OCT angiography . Biomed Opt Express . 2020 ; 11 : 330 – 345 . 32010520 10.1364/BOE.11.000330 PMC6968759

104. Guo Y , Hormel TT , Gao L , et al . Quantification of nonperfusion area in montaged widefield OCT angiography using deep learning in diabetic retinopathy . Ophthalmol Sci . 2021 ; 1 ( 2 ): 100027 . 36249293 10.1016/j.xops.2021.100027 PMC9560579

105. Hwang TS , Miao Z , Bhavsar K , et al . Visualization of 3 distinct retinal plexuses by projection-resolved optical coherence tomography angiography in diabetic retinopathy . JAMA Ophthalmol . 2016 ; 134 : 1411 – 1419 . 27812696 10.1001/jamaophthalmol.2016.4272 PMC5805384

106. Suh MH , Zangwill LM , Manalastas PIC , et al . Optical coherence tomography angiography vessel density in glaucomatous eyes with focal lamina cribrosa defects . Ophthalmology . 2016 ; 123 : 2309 – 2317 . 27592175 10.1016/j.ophtha.2016.07.023 PMC5360451

107. Yarmohammadi A , Zangwill LM , Diniz-Filho A , et al . Optical coherence tomography angiography vessel density in healthy, glaucoma suspect, and glaucoma eyes . Invest Ophthalmol Vis Sci . 2016 ; 57 ( 9 ): OCT451 – OCT459 . 27409505 10.1167/iovs.15-18944 PMC4968912

108. Chen HSL , Liu CH , Wu WC , Tseng HJ , Lee YS. Optical coherence tomography angiography of the superficial microvasculature in the macular and peripapillary areas in glaucomatous and healthy eyes . Invest Ophthalmol Vis Sci . 2017 ; 58 : 3637 – 3645 . 28728171 10.1167/iovs.17-21846

109. Jia Y , Wei E , Wang X , et al . Optical coherence tomography angiography of optic disc perfusion in glaucoma . Ophthalmology . 2014 ; 121 : 1322 – 1332 . 24629312 10.1016/j.ophtha.2014.01.021 PMC4082728

110. Simonett JM , Huang D , Jia Y. Quantitative optical coherence tomography angiography of the peripapillary circulation in glaucoma . Ann Eye Sci . 2017 ; 2 ( 2 ): 1 – 19 . 10.21037/aes.2017.01.03 PMC6112822 30167572

111. Nesper PL , Fawzi AA. New method for reducing artifactual flow deficits caused by compensation techniques in the choriocapillaris with optical coherence tomography angiography . Retina . 2022 ; 42 : 328 – 335 . 34653115 10.1097/IAE.0000000000003313 PMC8776592

112. Corvi F , Corradetti G , Tiosano L , McLaughlin JA , Lee TK , Sadda SR. Topography of choriocapillaris flow deficit predicts development of neovascularization or atrophy in age-related macular degeneration . Graefes Arch Clin Exp Ophthalmol . 2021 ; 259 : 2887 – 2895 . 33900443 10.1007/s00417-021-05167-3

113. Alagorie AR , Verma A , Nassisi M , et al . Quantitative assessment of choriocapillaris flow deficits surrounding choroidal neovascular membranes . Retina . 2020 ; 40 : 2106 – 2112 . 32658166 10.1097/IAE.0000000000002878

114. Shi Y , Zhang Q , Zhou H , et al . Correlations between choriocapillaris and choroidal measurements and the growth of geographic atrophy using swept source OCT imaging . Am J Ophthalmol . 2021 ; 224 : 321 – 331 . 33359715 10.1016/j.ajo.2020.12.015 PMC8058170

115. Moult EM , Shi Y , Zhang Q , et al . Analysis of correlations between local geographic atrophy growth rates and local OCT angiography-measured choriocapillaris flow deficits . Biomed Opt Express . 2021 ; 12 ( 7 ): 4573 . 34457433 10.1364/BOE.427819 PMC8367222

116. Moult EM , Alibhai AY , Lee BK , et al . A framework for multiscale quantitation of relationships between choriocapillaris flow impairment and geographic atrophy growth . Am J Ophthalmol . 2020 ; 214 : 172 – 187 . 31843474 10.1016/j.ajo.2019.12.006 PMC7951042

117. Greig EC , Moult EM , Despotovic IN , et al . Assessment of choriocapillaris flow prior to nascent geographic atrophy development using optical coherence tomography angiography . Invest Ophthalmol Vis Sci . 2024 ; 65 ( 1 ): 33 . 10.1167/iovs.65.1.33 PMC10807498 38236187

118. Gendelman I , Alibhai AY , Moult EM , et al . Topographic analysis of macular choriocapillaris flow deficits in diabetic retinopathy using swept-source optical coherence tomography angiography . Int J Retina Vitreous . 2020 ; 6 ( 1 ): 1 – 8 . 32206342 10.1186/s40942-020-00209-0 PMC7081691

119. Patel R , Wang J , Campbell JP , et al . Classification of choroidal neovascularization using projection-resolved optical coherence tomographic angiography . Invest Ophthalmol Vis Sci . 2018 ; 59 : 4285 – 4291 . 30372757 10.1167/iovs.18-24624 PMC6108780

120. Wang J , Hormel TT , Gao L , et al . Automated diagnosis and segmentation of choroidal neovascularization in OCT angiography using deep learning . Biomed Opt Express . 2020 ; 11 ( 2 ): 927 . 32133230 10.1364/BOE.379977 PMC7041469

121. Wang J , Hormel TT , Tsuboi K , et al . Deep learning for diagnosing and segmenting choroidal neovascularization in OCT angiography in a large real-world data set . Transl Vis Sci Technol . 2023 ; 12 ( 4 ): 15 . 10.1167/tvst.12.4.15 PMC10117225 37058103

122. Huang D , Jia Y , Rispoli M , Tan O , Lumbroso B. Optical coherence tomography angiography of time course of choroidal neovascularization in response to anti-angiogenic treatment . Retina . 2015 ; 35 : 2260 – 2264 . 26469535 10.1097/IAE.0000000000000846 PMC4627371

123. Spaide RF. Optical coherence tomography angiography signs of vascular abnormalization with antiangiogenic therapy for choroidal neovascularization . Am J Ophthalmol . 2015 ; 160 : 6 – 16 . 25887628 10.1016/j.ajo.2015.04.012

124. Muakkassa NW , Chin AT , De Carlo T , et al . Characterizing the effect of anti-vascular endothelial growth factor therapy on treatment-naive choroidal neovascularization using optical coherence tomography angiography . Retina . 2015 ; 35 : 2252 – 2259 . 26457400 10.1097/IAE.0000000000000836

125. Schreur V , Domanian A , Liefers B , et al . Morphological and topographical appearance of microaneurysms on optical coherence tomography angiography . Br J Ophthalmol . 2019 ; 103 : 630 – 635 . 10.1136/bjophthalmol-2018-312258 29925511

126. Parravano M , De Geronimo D , Scarinci F , et al . Diabetic microaneurysms internal reflectivity on spectral-domain optical coherence tomography and optical coherence tomography angiography detection . Am J Ophthalmol . 2017 ; 179 : 90 – 96 . 28483496 10.1016/j.ajo.2017.04.021

127. Parravano M , De Geronimo D , Scarinci F , et al . Progression of diabetic microaneurysms according to the internal reflectivity on structural optical coherence tomography and visibility on optical coherence tomography angiography . Am J Ophthalmol . 2019 ; 198 : 8 – 16 . 30308201 10.1016/j.ajo.2018.09.031

128. Gao M , Hormel TT , Guo Y , et al . Perfused and Nonperfused Microaneurysms Identified and Characterized by Structural and Angiographic OCT . Ophthalmol Retina . 2024 ; 8 : 108 – 115 . 37673397 10.1016/j.oret.2023.08.019

129. Gao M , Hormel T , Guo Y , et al . Identification and characterization of microaneurysm in OCT and OCT angiography . Invest Ophthalmol Vis Sci . 2022 ; 63 ( 7 ): 2899 – F0052 .

130. Huber KL , Stino H , Schlegl T , et al . Microaneurysm detection using high-speed megahertz optical coherence tomography angiography in advanced diabetic retinopathy . Acta Ophthalmol . 2024 ; 102 ( 5 ): e687 – e695 . 38126128 10.1111/aos.16619

131. Stattin M , Haas AM , Ahmed D , et al . Detection rate of diabetic macular microaneurysms comparing dye-based angiography and optical coherence tomography angiography . Sci Rep . 2020 ; 10 : 1 – 8 . 33005009 10.1038/s41598-020-73516-z PMC7530679

132. Salz DA , De Carlo TE , Adhi M , et al . Select features of diabetic retinopathy on swept-source optical coherence tomographic angiography compared with fluorescein angiography and normal eyes . JAMA Ophthalmol . 2016 ; 134 : 644 – 650 . 27055248 10.1001/jamaophthalmol.2016.0600 PMC5312730

133. Russell JF , Flynn HW , Sridhar J , et al . Distribution of diabetic neovascularization on ultra-widefield fluorescein angiography and on simulated widefield OCT angiography . Am J Ophthalmol . 2019 ; 207 : 110 – 120 . 31194952 10.1016/j.ajo.2019.05.031

134. Tsuboi K , Sasajima H , Kamei M. Collateral vessels in branch retinal vein occlusion: anatomic and functional analyses by optical coherence tomography angiography . Ophthalmol Retina . 2019 ; 3 : 767 – 776 . 31167729 10.1016/j.oret.2019.04.015

135. Suzuki N , Hirano Y , Yoshida M , et al . Microvascular abnormalities on optical coherence tomography angiography in macular edema associated with branch retinal vein occlusion . Am J Ophthalmol . 2016 ; 161 : 126 – 132.e1 . 26454243 10.1016/j.ajo.2015.09.038

136. Freund KB , Sarraf D , Leong BCS , Garrity ST , Vupparaboina KK , Dansingani KK. Association of optical coherence tomography angiography of collaterals in retinal vein occlusion with major venous outflow through the deep vascular complex . JAMA Ophthalmol . 2018 ; 136 : 1262 – 1270 . 30352115 10.1001/jamaophthalmol.2018.3586 PMC6248171

137. Schaal KB , Munk MR , Wyssmueller I , Berger LE , Zinkernagel MS , Wolf S. Vascular abnormalities in diabetic retinopathy assessed with swept-source optical coherence tomography angiography widefield imaging . Retina . 2019 ; 39 : 79 – 87 . 29135803 10.1097/IAE.0000000000001938

138. Sorour OA , Mehta N , Baumal CR , et al . Morphological changes in intraretinal microvascular abnormalities after anti-VEGF therapy visualized on optical coherence tomography angiography . Eye Vision . 2020 ; 7 : 1 – 9 . 32514410 10.1186/s40662-020-00195-2 PMC7262762

139. Arya M , Sorour O , Chaudhri J , et al . Distinguishing intraretinal microvascular abnormalities from retinal neovascularization using optical coherence tomography angiography . Retina . 2020 ; 40 : 1686 – 1695 . 31613839 10.1097/IAE.0000000000002671

140. Alam M , Le D , Son T , Lim JI , Yao X. AV-Net: deep learning for fully automated artery-vein classification in optical coherence tomography angiography . Biomed Opt Express . 2020 ; 11 : 5249 . 33014612 10.1364/BOE.399514 PMC7510886

141. Gao M , Guo Y , Hormel TT , et al . A deep learning network for classifying arteries and veins in montaged widefield OCT angiograms . Ophthalmol Sci . 2022 ; 2 ( 2 ): 100149 . 36278031 10.1016/j.xops.2022.100149 PMC9562370

142. Abtahi M , Le D , Lim JI , Yao X. MF-AV-Net: an open-source deep learning network with multimodal fusion options for artery-vein segmentation in OCT angiography . Biomed Opt Express . 2022 ; 13 : 4870 . 36187235 10.1364/BOE.468483 PMC9484445

143. Abtahi M , Le D , Ebrahimi B , Dadzie AK , Lim JI , Yao X. An open-source deep learning network AVA-Net for arterial-venous area segmentation in optical coherence tomography angiography . Commun Med . 2023 ; 3 : 54 . 37069396 10.1038/s43856-023-00287-9 PMC10110614

144. Murata T , Hirano T , Mizobe H , Toba S. OCT-angiography based artificial intelligence-inferred fluorescein angiography for leakage detection in retina [Invited] . Biomed Opt Express . 2023 ; 14 : 5851 . 38021144 10.1364/BOE.506467 PMC10659810

145. Le D , Alam M , Yao CK , et al . Transfer learning for automated octa detection of diabetic retinopathy . Transl Vis Sci Technol . 2020 ; 9 ( 2 ): 1 – 9 . 10.1167/tvst.9.2.35 PMC7424949 32855839

146. Heisler M , Karst S , Lo J , et al . Ensemble deep learning for diabetic retinopathy detection using optical coherence tomography angiography . Transl Vis Sci Technol . 2020 ; 9 ( 2 ): 1 – 11 . 10.1167/tvst.9.2.20 PMC7396168 32818081

147. Zang P , Gao L , Hormel TT , et al . DcardNet: diabetic retinopathy classification at multiple depths based on structural and angiographic optical coherence tomography . IEEE Trans Biomed Eng . 2021 ; 68 : 1859 – 1870 . 32986541 10.1109/TBME.2020.3027231 PMC8191487

148. Lo J , Yu TT , Ma D , et al . Federated learning for microvasculature segmentation and diabetic retinopathy classification of OCT data . Ophthalmol Sci . 2021 ; 1 ( 4 ): 100069 . 36246944 10.1016/j.xops.2021.100069 PMC9559956

149. Matten P , Scherer J , Schlegl T , et al . Multiple instance learning based classification of diabetic retinopathy in weakly-labeled widefield OCTA en face images . Sci Rep . 2023 ; 13 ( 1 ): 8713 . 37248309 10.1038/s41598-023-35713-4 PMC10226980

150. Liu Z , Wang C , Cai X , Jiang H , Wang J. Discrimination of diabetic retinopathy from optical coherence tomography angiography images using machine learning methods . IEEE Access . 2021 ; 9 : 51689 – 51694 .

151. Yang D , Tang Z , Ran A , et al . Assessment of parafoveal diabetic macular ischemia on optical coherence tomography angiography images to predict diabetic retinal disease progression and visual acuity deterioration . JAMA Ophthalmol . 2023 ; 141 : 641 – 649 . 37227703 10.1001/jamaophthalmol.2023.1821 PMC10214181

152. Lee YJ , Sun S , Kim YK , Jeoung JW , Park KH. Diagnostic ability of macular microvasculature with swept-source OCT angiography for highly myopic glaucoma using deep learning . Sci Rep . 2023 ; 13 ( 1 ): 5209 . 36997639 10.1038/s41598-023-32164-9 PMC10063664

153. Zang P , Hormel TT , Hwang TS , Bailey ST , Huang D , Jia Y. Deep-learning–aided diagnosis of diabetic retinopathy, age-related macular degeneration, and glaucoma based on structural and angiographic OCT . Ophthalmol Sci . 2023 ; 3 ( 1 ): 100245 . 36579336 10.1016/j.xops.2022.100245 PMC9791595

154. Huang X , Islam MR , Akter S , et al . Artificial intelligence in glaucoma: opportunities, challenges, and future directions . Biomed Eng Online . 2023 ; 22 : 126 . 38102597 10.1186/s12938-023-01187-8 PMC10725017

155. Akpinar MH , Sengur A , Faust O , Tong L , Molinari F , Acharya UR. Artificial intelligence in retinal screening using OCT images: A review of the last decade (2013–2023) . Comput Methods Programs Biomed . 2024 ; 254 : 108253 . 38861878 10.1016/j.cmpb.2024.108253

156. Schmidt-Erfurth U , Sadeghipour A , Gerendas BS , Waldstein SM , Bogunović H. Artificial intelligence in retina . Prog Retin Eye Res . 2018 ; 67 ( May ): 1 – 29 . 30076935 10.1016/j.preteyeres.2018.07.004

157. Guo Y , Hormel TT , Xiong H , Wang J , Hwang TS , Jia Y. Automated segmentation of retinal fluid volumes from structural and angiographic optical coherence tomography using deep learning . Transl Vis Sci Technol . 2020 ; 9 ( 2 ): 1 – 12 . 10.1167/tvst.9.2.54 PMC7552937 33110708

158. Zang P , Hormel TT , Wang J , et al . Interpretable diabetic retinopathy diagnosis based on biomarker activation map . IEEE Trans Biomed Eng . 2023 ; 71 : 14 – 25 . 37405891 10.1109/TBME.2023.3290541 PMC10796196