BMC pulmonary medicine · 2019
Robot-assisted bronchoscopy for pulmonary lesion diagnosis: results from the initial multicenter experience.
СтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Chaddha Udit, Kovacs Stephen P, Manley Christopher, Hogarth D Kyle, Cumbo-Nacheli Gustavo, Bhavani Sivasubramanium V, Kumar Rohit, Shende Manisha, Egan John P, Murgu Septimiu
Аннотация
BACKGROUND: The Robotic Endoscopic System (Auris Health, Inc., Redwood City, CA) has the potential to overcome several limitations of contemporary guided-bronchoscopic technologies for the diagnosis of lung lesions. Our objective is to report on the initial post-marketing feasibility, safety and diagnostic yield of this technology.
METHODS: We retrospectively reviewed data on consecutive cases in which robot-assisted bronchoscopy was used to sample lung lesions at four centers in the US (academic and community) from June 15th, 2018 to December 15th, 2018.
RESULTS: One hundred and sixty-seven lesions in 165 patients were included in the analysis, with an average follow-up of 185 ± 55 days. The average size of target lesions was 25.0 ± 15.0 mm. Seventy-one percent were located in the peripheral third of the lung. Pneumothorax and airway bleeding occurred in 3.6 and 2.4% cases, respectively. Navigation was successful in 88.6% of cases. Tissue samples were successfully obtained in 98.8%. The diagnostic yield estimates ranged from 69.1 to 77% assuming the cases of biopsy-proven inflammation without any follow-up information (N = 13) were non-diagnostic and diagnostic, respectively. The yield was 81.5, 71.7 and 26.9% for concentric, eccentric and absent r-EBUS views, respectively. Diagnostic yield was not affected by lesion size, density, lobar location or centrality.
CONCLUSIONS: RAB implementation in community and academic centers is safe and feasible, with an initial diagnostic yield of 69.1-77% in patients with lung lesions that require diagnostic bronchoscopy. Comparative trials with the existing bronchoscopic technologies are needed to determine cost-effectiveness of this technology.
METHODS: We retrospectively reviewed data on consecutive cases in which robot-assisted bronchoscopy was used to sample lung lesions at four centers in the US (academic and community) from June 15th, 2018 to December 15th, 2018.
RESULTS: One hundred and sixty-seven lesions in 165 patients were included in the analysis, with an average follow-up of 185 ± 55 days. The average size of target lesions was 25.0 ± 15.0 mm. Seventy-one percent were located in the peripheral third of the lung. Pneumothorax and airway bleeding occurred in 3.6 and 2.4% cases, respectively. Navigation was successful in 88.6% of cases. Tissue samples were successfully obtained in 98.8%. The diagnostic yield estimates ranged from 69.1 to 77% assuming the cases of biopsy-proven inflammation without any follow-up information (N = 13) were non-diagnostic and diagnostic, respectively. The yield was 81.5, 71.7 and 26.9% for concentric, eccentric and absent r-EBUS views, respectively. Diagnostic yield was not affected by lesion size, density, lobar location or centrality.
CONCLUSIONS: RAB implementation in community and academic centers is safe and feasible, with an initial diagnostic yield of 69.1-77% in patients with lung lesions that require diagnostic bronchoscopy. Comparative trials with the existing bronchoscopic technologies are needed to determine cost-effectiveness of this technology.