MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcv.2023.105521

Медицинская библиотека · Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology · 2023

Classification of "Near-patient" and "Point-of-Care" SARS-CoV-2 Nucleic Acid Amplification Test Systems and a first approach to evaluate their analytical independence of operator activities.

Buchta Christoph, Zeichhardt Heinz, Badrick Tony, Coucke Wim, Wojtalewicz Nathalie, Griesmacher Andrea, Aberle Stephan W, Schellenberg Ingo, Jacobs Ellis, Nordin Gunnar, Schweiger Christian, Schwenoha Karin, Luppa Peter B, Gassner Ulrich M, Wagner Thomas, Kammel Martin

СтатьяПеревод в очереди

Аннотация

BACKGROUND

European legislation defines as "near-patient testing" (NPT) what is popularly and in other legislations specified as "point-of-care testing" (POCT). Systems intended for NPT/POCT use must be characterized by independence from operator activities during the analytic procedure. However, tools for evaluating this are lacking. We hypothesized that the variability of measurement results obtained from identical samples with a larger number of identical devices by different operators, expressed as the method-specific reproducibility of measurement results reported in External Quality Assessment (EQA) schemes, is an indicator for this characteristic.

MATERIALS AND

METHODS

Legal frameworks in the EU, the USA and Australia were evaluated about their requirements for NPT/POCT. EQA reproducibility of seven SARS-CoV-2-NAAT systems, all but one designated as "POCT", was calculated from variabilities in Ct values obtained from the respective device types in three different EQA schemes for virus genome detection.

RESULTS

A matrix for characterizing test systems based on their technical complexity and the required operator competence was derived from requirements of the European In Vitro Diagnostic Regulation (IVDR) 2017/746. Good EQA reproducibility of the measurement results of the test systems investigated implies that different users in different locations have no recognizable influence on their measurement results.

CONCLUSION

The fundamental suitability of test systems for NPT/POCT use according to IVDR can be easily verified using the evaluation matrix presented. EQA reproducibility is a specific characteristic indicating independence from operator activities of NPT/POCT assays. EQA reproducibility of other systems than those investigated here remains to be determined.

Для цитирования: Buchta Christoph, Zeichhardt Heinz, Badrick Tony, Coucke Wim, Wojtalewicz Nathalie, Griesmacher Andrea, Aberle Stephan W, Schellenberg Ingo, Jacobs Ellis, Nordin Gunnar, Schweiger Christian, Schwenoha Karin, Luppa Peter B, Gassner Ulrich M, Wagner Thomas, Kammel Martin. Classification of "Near-patient" and "Point-of-Care" SARS-CoV-2 Nucleic Acid Amplification Test Systems and a first approach to evaluate their analytical independence of operator activities.. Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology. 2023. DOI: 10.1016/j.jcv.2023.105521.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.