Journal of immunological methods · 2017

Laboratory-based performance evaluation of PIMA CD4+ T-lymphocyte count point-of-care by lay-counselors in Kenya.

Сравнительное исследованиеОценочное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Zeh Clement, Rose Charles E, Inzaule Seth, Desai Mitesh A, Otieno Fredrick, Humwa Felix, Akoth Benta, Omolo Paul, Chen Robert T, Kebede Yenew, Samandari Taraz

Аннотация

BACKGROUND: CD4+ T-lymphocyte count testing at the point-of-care (POC) may improve linkage to care of persons diagnosed with HIV-1 infection, but the accuracy of POC devices when operated by lay-counselors in the era of task-shifting is unknown. We examined the accuracy of Alere's Pima™ POC device on both capillary and venous blood when performed by lay-counselors and laboratory technicians.

METHODS: In Phase I, we compared the perfomance of POC against FACSCalibur™ for 280 venous specimens by laboratory technicians. In Phase II we compared POC performance by lay-counselors versus laboratory technicians using 147 paired capillary and venous specimens, and compared these to FACSCalibur™. Statistical analyses included Bland-Altman analyses, concordance correlation coefficient, sensitivity, and specificity at treatment eligibility thresholds of 200, 350, and 500cells/μl.

RESULTS: Phase I: POC sensitivity and specificity were 93.0% and 84.1% at 500cells/μl, respectively. Phase II: Good agreement was observed for venous POC results from both lay-counselors (concordance correlation coefficient (CCC)=0.873, bias -86.4cells/μl) and laboratory technicians (CCC=0.920, bias -65.7cells/μl). Capillary POC had good correlation: lay-counselors (CCC=0.902, bias -71.2cells/μl), laboratory technicians (CCC=0.918, bias -63.0cells/μl). Misclassification at the 500 cells/μl threshold for venous blood was 13.6% and 10.2% for lay-counselors and laboratory technicians and 12.2% for capillary blood in both groups. POC tended to under-classify the CD4 values with increasingly negative bias at higher CD4 values.

CONCLUSIONS: Pima™ results were comparable to FACSCalibur™ for both venous and capillary specimens when operated by lay-counselors. POC CD4 testing has the potential to improve linkage to HIV care without burdening laboratory technicians in resource-limited settings.