Performance Evaluation of a Simple Strategy to Optimize Formula Constants for Zero Mean or Minimal Standard Deviation or Root-Mean-Squared Prediction Error in Intraocular Lens Power Calculation.

Langenbucher Achim, Szentmáry Nóra, Wendelstein Jascha, Cayless Alan, Hoffmann Peter, Gatinel Damien

СтатьяОценочное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To investigate the performance of a simple prediction scheme for the formula constants optimized for a mean (MPE), standard deviation (SDPE) or root-mean-squared refractive prediction error (RMSPE).

DESIGN

Retrospective cross-sectional study.

METHODS

Using IOLMaster 700 biometric data from 888 eyes treated with the Hoya Vivinex lens and 821 eyes treated with the Alcon SA60AT lens, plus the power of the implanted lens and postoperative spherical equivalent refraction, optimized constants for SRKT, Hoffer Q, Holladay 1, Haigis, and K6 formulae were calculated using an iterative nonlinear optimization for zero MPE and minimal SDPE and RMSPE. Start values were detuned by ±1.5 from the MPE optimized constants and formula constants generated using the simple prediction scheme were compared to the corresponding directly optimized constants.

RESULTS

For all 5 formulae under test and with both datasets, constants optimized using the simple scheme showed excellent agreement with those from the iterative method with either MPE or RMSPE used as the optimization metric and good agreement with SDPE as the metric. Constants optimized for zero MPE or minimal RMSPE agreed within 0.05, whereas constants for minimal SDPE could be systematically off by up to 0.6 from the MPE values, making SDPE unsuitable as an optimization metric.

CONCLUSIONS

This simple formula constant optimization scheme performs excellently for 4 disclosed formulae and one nondisclosed formula in our 2 monocentric datasets with zero MPE or minimal RMSPE as metrics. Multicentric studies with other study populations and biometers are required to further investigate the clinical applicability.

Для цитирования: Langenbucher Achim, Szentmáry Nóra, Wendelstein Jascha, Cayless Alan, Hoffmann Peter, Gatinel Damien. Performance Evaluation of a Simple Strategy to Optimize Formula Constants for Zero Mean or Minimal Standard Deviation or Root-Mean-Squared Prediction Error in Intraocular Lens Power Calculation.. American journal of ophthalmology. 2024. DOI: 10.1016/j.ajo.2024.08.043.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.