Медицинская библиотека · American journal of rhinology · 2003

Ventricular tachycardia as an electrocardiography artifact during functional endoscopic sinus surgery.

Gaiser Robert R, Demetry Daphne J, Schlosser Rodney

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Functional endoscopic sinus surgery is frequently performed during conscious sedation not provided by an anesthesiologist. Several monitors are used during conscious sedation, one of which is the electrocardiogram (EKG). Equipment used in the operating room may interfere with the EKG tracing.

METHODS

A 30-year-old man underwent his fourth functional endoscopic sinus surgery during general anesthesia. For the procedure, a Xomed Straight Shot microdebrider was used.

RESULTS

Because of stray electrical current, the use of the microdebrider caused the EKG waveform to resemble ventricular tachycardia.

CONCLUSION

This case alerts the clinician to this possible occurrence to prevent inappropriate therapeutics from being initiated.

Для цитирования: Gaiser Robert R, Demetry Daphne J, Schlosser Rodney. Ventricular tachycardia as an electrocardiography artifact during functional endoscopic sinus surgery.. American journal of rhinology. 2003.