Can automated haematology analysers discriminate thalassaemia from iron deficiency?

Marti H R, Fischer S, Killer D, Bürgi W

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

The use of automated analysers in population screening for beta-thalassaemia has been a matter of controversy. The new fully automated haematology analyser Sysmex E-5000 (Toa Medical Electronics Co. Ltd) facilitates the discrimination of heterozygous thalassaemia from iron deficiency anaemia. In addition to haemoglobin, mean corpuscular haemoglobin and mean corpuscular volume, the red cell size-distribution width is measured. In patients with hypochromic microcytic red cells, the Sysmex data have been evaluated and compared with the indices described by England and Fraser [Lancet i, pp. 449-452, 1973], Mentzer [Lancet i, p. 882, 1973] and by Shine and Lal [Lancet i, pp. 692-694, 1977]. For the detection of beta-thalassaemia trait, the size-distribution width is superior to the previously described indices. The sensitivity is 79%, the specificity 95% and the predictive value for a positive test 94%.

Для цитирования: Marti H R, Fischer S, Killer D, Bürgi W. Can automated haematology analysers discriminate thalassaemia from iron deficiency?. Acta haematologica. 1987. DOI: 10.1159/000205871.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.