MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1034/j.1399-6576.2002.460617.x

Медицинская библиотека · Acta anaesthesiologica Scandinavica · 2002

Monitoring functional residual capacity (FRC) by quantifying oxygen/carbon dioxide fluxes during a short apnea.

Stenqvist Ola, Olegård C, Søndergaard S, Odenstedt H, Kárason S, Lundin S

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Clinically applicable methods for measuring FRC are currently lacking. This study presents a new method for FRC monitoring based on quantification of metabolic gas fluxes of O2 and CO2 during a short apnea.

METHODS

Base line exchange of oxygen and carbon dioxide was measured with indirect calorimetry. End-tidal ( approximately alveolar) O2 and CO2 concentrations were measured before and after a short apnea, 8-12 s, and FRC was calculated according to standard washin/washout formulas taking into account the increased solubility of CO2 in blood when the tension is increased during the apnea. The method was tested in a lung model with CO2 excretion and O2 consumption achieved by combustion of hydrogen and implemented in six ventilator-treated patients with acute respiratory failure (ARF).

RESULTS

In the lung model the method showed excellent correlation (r = 0.98) with minimal bias (34 ml) and a good precision, limits of agreement being 160 and -230 ml, respectively, compared to the reference method. In six ARF patients changes in FRC induced by increase or decrease in PEEP and measured with the O2/CO2 flux FRC method corresponded well with changes in reference values of FRC (r = 0.76-0.94).

CONCLUSIONS

A new method has been proposed in which FRC could be monitored from measurements of physiological fluxes of gases during a short apnea with the use of standard ICU equipment and some calculations. We anticipate that with further development, this technique could provide a new tool for monitoring respiratory changes and ventilator management in the ICU.

Для цитирования: Stenqvist Ola, Olegård C, Søndergaard S, Odenstedt H, Kárason S, Lundin S. Monitoring functional residual capacity (FRC) by quantifying oxygen/carbon dioxide fluxes during a short apnea.. Acta anaesthesiologica Scandinavica. 2002. DOI: 10.1034/j.1399-6576.2002.460617.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.