Quantitative OCT-based longitudinal evaluation of intracorneal ring segment implantation in keratoconus.

Pérez-Merino Pablo, Ortiz Sergio, Alejandre Nicolas, Jiménez-Alfaro Ignacio, Marcos Susana

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To characterize the geometrical properties of keratoconic corneas upon intracorneal ring segments (ICRS) implantation, using custom-developed optical coherence tomography (OCT).

METHODS

Ten keratoconic corneas were measured pre- and post-ICRS surgery (7, 30, and 90 days). Corneal topographic and pachymetric maps were obtained from three-dimensional (3D) images acquired with OCT, provided with custom algorithms for image analysis, distortion correction, and quantification. The 3D positioning of the ICRS was also estimated longitudinally, relative to the pupil center and iris plane.

RESULTS

Preoperatively, the average corneal radii of curvature were 7.02 ± 0.54 mm (anterior) and 5.40 ± 0.77 mm (posterior), and the minimum corneal thickness was 384 ± 60 μm. At 90 days, the average corneal radii of curvature were 7.26 ± 0.53 mm (anterior) and 5.44 ± 0.71 mm (posterior), and the minimum corneal thickness was 396 ± 46 μm. ICRS implantation produced a significant decrease of corneal power (by 1.71 ± 1.83 diopters [D] at 90 days). Corneal irregularities (defined by high order Zernike terms of the corneal elevation maps) and the corneal thickness distribution decreased in some patients and increased in others. The 3D ICRS depth matched the planned ICRS depth well (within 23.93 ± 23.49 μm). On average, ICRS showed an overall tilt of -6.8 ± 2.6° (temporal) and -2.1 ± 0.8° (superior) at 7 days.

CONCLUSIONS

Spectral OCT (sOCT) provided with distortion correction and analysis tools, is an excellent instrument for evaluating the changes produced by ICRS in keratoconic corneas, and for analyzing the 3D ICRS position during the follow up. ICRS produced flattening on the anterior corneal surface, although the benefit for corneal surface regularization varied across patients.

Для цитирования: Pérez-Merino Pablo, Ortiz Sergio, Alejandre Nicolas, Jiménez-Alfaro Ignacio, Marcos Susana. Quantitative OCT-based longitudinal evaluation of intracorneal ring segment implantation in keratoconus.. Investigative ophthalmology & visual science. 2013. DOI: 10.1167/iovs.13-12401.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.