DOI 10.1111/ceo.12113

Медицинская библиотека · Clinical & experimental ophthalmology · 2013

Corneal thickness and elevation measurements using swept-source optical coherence tomography and slit scanning topography in normal and keratoconic eyes.

Jhanji Vishal, Yang Bingzhi, Yu Marco, Ye Cong, Leung Christopher K S

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

To compare corneal thickness and corneal elevation using swept source optical coherence tomography and slit scanning topography.

DESIGN

Prospective study.

PARTICIPANTS

41 normal and 46 keratoconus subjects.

METHODS

All eyes were imaged using swept source optical coherence tomography and slit scanning tomography during the same visit. Mean corneal thickness and best-fit sphere measurements were compared between the instruments.

MAIN OUTCOME MEASURES: Agreement of measurements between swept source optical coherence tomography and scanning slit topography was analyzed. Intra-rater reproducibility coefficient and intraclass correlation coefficient were evaluated.

RESULTS

In normal eyes, central corneal thickness measured by swept source optical coherence tomography was thinner compared with slit scanning topography (p < 0.0001) and ultrasound pachymetry (p = < .0001). Ultrasound pachymetry readings had better 95% limits of agreement with swept source optical coherence tomography than slit scanning topography. In keratoconus eyes, central corneal thickness was thinner on swept source optical coherence tomography than slit scanning topography (p = 0.081) and ultrasound pachymetry (p = 0.001). There were significant differences between thinnest corneal thickness, and, anterior and posterior best-fit sphere measurements between both instruments (p < 0.05 for all). Overall, reproducibility coefficients and intraclass correlation coefficients were significantly better with swept source optical coherence tomography for measurement of central corneal thickness, anterior best-fit sphere and, posterior best-fit sphere (all p < 0.001).

CONCLUSIONS

Corneal thickness and elevation measurements were significantly different between swept source optical coherence tomography and slit scanning topography. With better reproducibility coefficients and intraclass correlation coefficients, swept source optical coherence tomography may provide a reliable alternative for measurement of corneal parameters.

Для цитирования: Jhanji Vishal, Yang Bingzhi, Yu Marco, Ye Cong, Leung Christopher K S. Corneal thickness and elevation measurements using swept-source optical coherence tomography and slit scanning topography in normal and keratoconic eyes.. Clinical & experimental ophthalmology. 2013. DOI: 10.1111/ceo.12113.