DOI 10.1097/icl.0b013e3182657da5
The effect of optic asphericity on visual rehabilitation of corneal ectasia with a prosthetic device.
Аннотация
OBJECTIVES
To study the effect of optic asphericity on visual rehabilitation of corneal ectasia with a prosthetic device.
METHODS
Subjects with corneal ectasia treated with a fluid-ventilated gas-permeable prosthetic device of diameter 18.0-19.0 mm, who had subjective improvement of good Snellen vision with introduction of optic asphericity, were studied. Best corrected Snellen visual acuity (BCVA) under standard illumination, high contrast visual acuity (HCVA), low contrast visual acuity (LCVA), and wavefront aberrations were measured in a sequence of devices that varied per patient only in presence or amount of ellipsoidal front surface optical eccentricity (FSE).
RESULTS
Five eyes of 5 subjects were studied. (M:F = 3:2; Age: 20-76). Mean steepest SimK was 57.72±8.30 D. BCVA was ≥20/30 in all eyes in all prosthetic devices, regardless of FSE. Although FSE improved BCVA, HCVA, and LCVA in each patient, no optimal amount could be identified in this small series. Asphericity in the form of 0.6 or 0.8 FSE improved HCVA, LCVA, or both in each patient. FSE was associated with a trend toward reduction of higher-order aberrations, particularly coma.
CONCLUSIONS
Optic asphericity shows promise for optimization of vision in the rehabilitation of corneal ectasia with a prosthetic device.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.