Measurement of the speed of sound in follicular fluid.

Gooding M J, Barber D, Kennedy S H, Noble J A

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Measurement of ovarian follicles by ultrasound is common practice in fertility treatment. However, the effect of the speed of sound is not taken into account. We present results from a study aimed at measuring this.

METHODS

The speed of sound was measured in samples of follicular fluid aspirated from patients undergoing fertility treatment. The transmitted and received pulses from a single element ultrasound transducer were recorded using a digital oscilloscope for a pulse passed through a sample of the fluid. The distance over which the pulse travelled was known from calibration with pure water. Variation with temperature was investigated in the range 25-45 degrees C. Dependence on ultrasound frequency, patient and time from aspiration were also investigated.

RESULTS

The speed of sound in follicular fluid was found to be 1550+/-3 m/s at 37.3 degrees C using 5.0 MHz ultrasound. The speed varied from 1528+/-3 m/s at 24.8 degrees C to 1561+/-3 m/s at 44.8 degrees C. Variation with patient, time and frequency were not detected.

CONCLUSION

The speed of sound in follicular fluid at body temperature is 1550 m/s. This small difference from the speed assumed by the ultrasound machine corresponds to the systematic bias in volume measurement evident in previously published results.

Для цитирования: Gooding M J, Barber D, Kennedy S H, Noble J A. Measurement of the speed of sound in follicular fluid.. Human reproduction (Oxford, England). 2004. DOI: 10.1093/humrep/deh591.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.